![Molift QUICK RAISER Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/molift/quick-raiser/quick-raiser_operation-manual_1824207016.webp)
Page 16 of 0
E Rev. G/ 09/006
Maintenance
Daily maintenance
Detergents must be pH-neutral. Do not use
solvents or strong liquids - this may damage
surfaces on the lifter. For disinfection when
needed; use isopropyl alcohol. Avoid abra-
sive cleaning products.
Clean surfaces with a damp cloth using an appropri-
ate detergent.
Monthly maintenance
• Remove hair and pile from the castors and verify
that the castors rotates without abnormal friction
• Verify that the contacts of the hand control and
charger is firmly in place
• Clean contact and hand control with spirit or
similar to remove grease and dirt
• Verify that the cables for the hand control, actua-
tor and charger are intact
• Run the lifter up and down to verify that it runs
normally and without any abnormal noise
Fault finding
If you have access to more than one lifter, it may be
helpful to change parts across lifters to help identify
any faults. For instance, if you suspect the fault to
be inside the column, exchange the column with at
functional one from the other lifter to see if the prob-
lem is fixed.
Across you will find a table with fault finding proce
-
dures to identify and correct faults on the lifter:
Symptom
Possible cause/
action
Lifting column is
wobbly
Locking bolts are not prop-
erly secured or lifting column
is not properly in place in the
chassis /
check if the col-
umn is correctly placed in the
chassis and/or tighten bolts
Lifter only moves
one way (up
or
down) / legs only
moves one way
(in
or
out)
Fault on hand control /
ex-
change with alterative control
from other lifter and/or new
hand control
One of the relays / electronics
has failed/ exchange the con-
trol unit/relays
/ contact your
local service representative
The lifter moves
up and/or down
by itself / legs
move by them-
selves
Short circuit between the
hand control and lifter /
clean
the connectors of the hand
control and chassis with spir-
its to remove grease and dirt
The hand control has failed/
exchange hand control
/ con-
tact your local service repre-
sentative
Relays/electronics malfunc-
tions /
exchange electron-
ics/relays / contact your local
service representative
The lifting arm
does not move
/ legs does not
move
Discharged battery /
ex-
change battery with another
or charge battery
Bad or wobbly connection
from lifting column to chas-
sis /
clean contacts in column
base and inside chassis - ad-
just and tighten column bolts
Emergency stop activated /
turn clock wise to reset
Hand control disconnected /
reconnect hand control
The hand control has failed/
exchange hand control
/ con-
tact your local service repre-
sentative
Relays/electronics malfunc-
tions /
exchange electron-
ics/relays / contact your local
service representative
Motor has failed /
exchange
motor / contact your local
service representative