Maintenance and Troubleshooting
5023399 A
19
2.
Set up a protocol through the
Acquire > New Protocol
menu
command, and then on the
Mode/Rate
tab, select the
Gap-free
mode. On the
Outputs
tab, for each of the
Analog OUT
Channels
, select the matching signal (for example,
OUT 1
) from
its list.
3.
Attach the digitizer’s outputs to an oscilloscope or 10-bit digital
volt meter (DVM).
4.
For each output signal, use the software’s
Real Time Control
panel to change voltage levels. For analog outputs, either use the
spinners, or type a value and press the
Enter
key. For digital
outputs, click on the box corresponding to the digital bit to be
tested. Verify the output signal levels on the oscilloscope or
DVM.