Using the Trueprep

 

 AUTO v2 

®

7.1 

Charging/Using the Trueprep

 

 AUTO v2 with the AC Adapter

®

7.1.1  Press the

 Power 

button to switch on the

 Trueprep

 

 AUTO v2

7.1.2  Low battery

 

 

 

If the battery charge is low, you will see the above alert message

 

 

The Charging/Low Battery LED Indicator will glow red and you will also hear a beeping 

® 

sound.  The 

Trueprep  AUTO v2

 device can not begin a  new run if the battery is low. 

Fig 4: Low battery LED glowing red

7.1.3  If the battery charge is full or sufficient to carry out a run, the Charging/Low Battery LED Indicator 

® 

will stay off. You may proceed to use

 Trueprep AUTO v2 

on battery charge.

® 

 

For charging or to use 

Trueprep AUTO v2

 on mains, connect the 

AC Adapter

 to the 

Charging 

Port

 on the right side of the back panel of the device.

BATTERY LOW !

CONNECT CHARGER 

®

®

Trueprep 

 AUTO 

v2

Содержание Trueprep AUTO v2

Страница 1: ...True AUTO prep v2 User Manual 603042001 REF IVD Universal Cartridge Based Sample Prep Device Trueprep AUTO v2 ...

Страница 2: ...provided with device 5 6 The Trueprep AUTO v2 5 6 1 A Top Panel View 5 6 1B Back Panel View 6 6 2 The Cartridge 6 7 Using the Trueprep AUTO v2 7 7 1 Charging Using the Trueprep AUTO v2 with the AC Adapter 7 7 2 Plugging in the reagent bottles 8 7 3 Switching ON OFF the Trueprep AUTO v2 11 7 4 Placing the Cartridge in the device and starting the run 11 8 Errors Alerts and Maintenance 14 8 1 Error M...

Страница 3: ...eagent Reset Card for 50Tests Pack size 10 Figure 10 The cartridge holder opened after eject button is pressed 11 Figure 11 Cartridge being placed in cartridge holder 12 Figure 12 Cartridge placed in cartridge holder in correct orientation 12 Figure 13 Replaceable tray in cartridge holder 13 Figure 14 Removal of Replaceable Tray from cartridge holder 13 Figure 15 Replacement of discarded Replaceab...

Страница 4: ...nd does not require highly skilled personnel to carry out the extraction process All the waste from processing of the sample is contained within the cartridge dump area thus posing no risk from potentially bio hazardousmaterial Thedevicehasauniversalprotocolandcanworkwithallkindsofsamplessuchassputum BAL whole blood serum plasma tissue stool urine CSF pleural fluid lymph node aspirate pus peritone...

Страница 5: ...condensing Theunitshouldbestoredonaflat drysurface 3 Specifications AUTOv2complieswiththeemissionandimmunityrequirementsdescribedinIEC61326 Trueprep series 4 Precautions CheckforLowBatterybeforeusingtheTrueprep AUTOv2 seesection7 1 Insertandremovethecartridgegentlyfromtheholder useofexcessiveforcecoulddamage it Donotspillliquidsinsidethecartridgeholder Do not perform an extraction in the presence ...

Страница 6: ...ith Trueprep AUTO AUTOv2UniversalCartridgeBasedSamplePrepKit REF60203AR25 REF60207AR50 consistingof UniversalReagentPack 25 50tests UniversalCartridgePack 25 50Nos Disposabletransferpipettegraduated 25x3ml 50x3ml Packageinsert 1No 6 TheTrueprep AUTOv2 6 1A Top PanelView Figure 1 The Trueprep AUTO v2 Top View with cartridge holder shown in the front panel ThePowerbuttonswitchesonandswitchesofftheTr...

Страница 7: ...luewhentheTrueprep AUTOv2isbeingcharged 6 1B BackPanelView Figure 2 The Trueprep AUTO v2 device Back Panel with Reagent Pack of the Trueprep AUTO v2 Universal Cartridge Based Sample Prep Kit 6 2 TheCartridge Samples are processed on disposable single use cartridges that are inserted into the cartridge holder These cartridges come pre loaded with Internal Positive Control and only require the addit...

Страница 8: ... and you will also hear a beeping sound TheTrueprep AUTOv2devicecannotbegina newrunifthebatteryislow Fig 4 Low battery LED glowing red 7 1 3 Ifthebatterychargeisfullorsufficienttocarryoutarun theCharging LowBatteryLEDIndicator willstayoff YoumayproceedtouseTrueprep AUTOv2onbatterycharge For charging or to use Trueprep AUTO v2 on mains connect the ACAdapter to the Charging Portontherightsideoftheba...

Страница 9: ...ACAdapterunlessitisbeingusedtopowerthedevice Figure 6 The Blue Charging LED Indicator Glowing 7 2 PluggingintheReagentBottlePackof25TorReagentPackof50T 7 2 1 PluggingintheReagentBottlePackof25Tpacksize 1 Beforestartingarunforthefirsttime pluginthecolourcodedreagentbottlesfromtheTrueprep AUTO Sample Prep Kit Reagent Pack into the back panel of the device matching the colour of the cap on the panel ...

Страница 10: ...ugging in the reagent pack of 50T pack size 1 Before starting a run for the first time plug in the reagent pack from the Trueprep AUTO v2 Sample Prep Kit to the backside of the device Figure 8 Discard the used reagent pack in a propermanner 2 The device will prompt when 50 tests have been completed and the reagent pack have to be replacedbyfollowingbufferresettingprocedureasperPointNo 7 2 5 1or7 2...

Страница 11: ...l be displayed on screen alternatively once buffer count reaches 50 Press Eject insert Reagent Reset Card provided with the new Reagent pack close the tray and press start Once the buffer count has reset to zero following will be displayed on screen BY PRESSING START EJECT SW SHUTTING DOWN CHANGE REAGENT PACK RESET SHUTTING DOWN SETTINGS SAVED SHUTTING DOWN INSERT REAGENT RESET CARD START SHUTTING...

Страница 12: ...ingscreenwillbedisplayed releasethepowerbutton Thedevicewillturnoffautomaticallywithin30seconds 7 4 PlacingtheCartridgeinthedeviceandstartingtherun Important Onlyusethecartridgeprovidedinthecorrespondingsampleprepkits 7 4 1 PresstheEjectbuttontoopenthecartridgeholder Youwillbepromptedwiththemessage 11 TRUEPREP AUTO BUFFER COUNT SHUTTING DOWN INSERT CARTRIDGE Figure10 Thecartridgeholderopenedaftere...

Страница 13: ...idge holder to remain open andthecartridgewillnotbeinsertedinthedevice Figure 11 Cartridge being placed in cartridge holder 7 4 3 Push the cartridge holder gently inwards to close it shut Press the Start Button to proceed This is the start of the Trueprep AUTO v2 sample preparation process After reading the QR code of Cartridge Extraction process starts If the QR code is read then the same cartrid...

Страница 14: ...dge holder as shown in the Figure 14 Replaceable Step 2 Placement of the tray 2a Place the tray in cartridge holder as shown in Figure 15 Replaceable i Inspect the tray in the cartridge holder Replaceable ii If there are liquids spilled in the tray take out the tray and discard as per the instructions in Section8 3 iii Cleanthecartridgeholderwith70 IsopropylAlcohol IPA asdescribedinSection8 3 iv R...

Страница 15: ...on PressEjecttoexit alternatively Solution Contact Molbio s support team E1 CARTRIDGE VALVE ERROR Press Eject to exit E2 CARTRIDGE ERROR The Screen displays the error message E2 CARTRIDGE ERROR and instruction PressEjecttoexit alternatively Solution Contact Molbio s support team The Screen displays the error message E3 CARTRIDGECLOGGED and instruction PressEjecttoexit alternatively Solution Contac...

Страница 16: ...ate of cartridge The Screen displays the error message E6 CARTRIDGE NOT LOADED and instruction PressEjecttoexit alternatively Solution Contact Molbio s support team Press Eject to exit E6 CARTRIDGE NOT LOADED Press Eject to exit Reagent Pack Errors and Warnings E7 E8 E7 EXPIRED RESET CARD The Screen displays the error message E7 EXPIRED RESET CARD and instruction Press Eject to exit alternatively ...

Страница 17: ...W1 SBC NOT Responding Reagent Pack Errors and Warnings E11 E11 RTD L ERROR E12 The Screen displays the error message E12 RTD E ERROR Note This is for Elution Heater E11 RTD E ERROR Device Errors and warnings E11 The Screen displays the error message E11 RTD L ERROR Note This is for Lysis Heater Solution Contact Molbio s support team E11 RTD L ERROR E12 The Screen displays the error message E12 RTD...

Страница 18: ...ld Start Button and Power Button on the Trueprep AUTO v2 Firmware Version and Total run status counts will be displayed on the screen as given below Ÿ Release both the buttons when following is displayed Ÿ In themenu thefirstlineofLCDdisplayshowsdifferentfunctionoptions Thesecondlinedisplays navigating options for next Eject Button and back Start Button The Power Button allows selectionofcurrently...

Страница 19: ...e it is selected tray will be ejected Following screens will toggle with 2 seconds interval Ÿ Insert cartridge and press start the flushing process will begin Ÿ The cartridge will be ejected once flushing is completed following will be displayed on screen and device will go back to menu Ÿ Press Power button to exit without flushing EXIT OK EXIT OK EXIT OK CONNECT DISTILLED WATER Trueprep AUTO v2 E...

Страница 20: ...eshly prepared 0 5 Sodium Hypochlorite solutionandacleanclothortissue Thisshouldbedoneonceeverydayand orwhenthedeviceis movedtoadifferentworkspace 8 3 3Spills If you notice a spill in the tray placed in the cartridge holder dispose the tray Clean Replaceable the insides of the cartridge holder with clean cloth or tissue or swab dipped in 70 Isopropyl Alcohol IPA Waitfor5minutesbeforeinsertinganew ...

Страница 21: ...20 In Vitro Diagnostic Medical Device Trueprep AUTO v2 ...

Страница 22: ... FRAGILE KEEP DRY Fully automatic operation Rapid sample prepping Mains rechargeable battery powered AUTO v2 Universal Cartridge Based Sample Prep Kit Portable This Side Up Medical Electronic Rush Do Not Stack Fragile Keep Dry True prep Trueprep AUTO v2 ...

Страница 23: ... Registered Office Manufacturing Unit Plot No L 46 Phase II D Verna Industrial Estate Verna Goa 403 722 INDIA www molbiodiagnostics com Molbio Diagnostics Private Limited Qarad EC REP b v b a Pas 257 B 2440 Geel Belgium ...

Отзывы: