9
3. A set of bottles is sufficient for 25 tests.
4. The device will prompt when 25 tests have been completed and the bottles have to be replaced
by following manual buffer resetting procedure as per Point No. 7.2.5.1.
®
Figure 7: The
Trueprep AUTO v2
Back Panel, with tubing and caps for plugging in reagent bottles
®
from the Reagent Pack of 25T pack size of the
Trueprep AUTO
Universal Cartridge Based Sample Prep Kit
7.2.2 Plugging in the reagent pack of 50T pack size:
®
1. Before starting a run for the first time, plug in the reagent pack from the
Trueprep AUTO v2
Sample Prep Kit to the backside of the device (Figure 8). Discard the used reagent pack in a
proper manner.
2. The device will prompt when 50 tests have been completed and the reagent pack have to be
replaced by following buffer resetting procedure as per Point No. 7.2.5.1 or 7.2.5.2.
®
Figure 8: The
Trueprep AUTO v2
device
Back Panel with Reagent Pack (50T) of
®
the
Trueprep AUTO v2
Universal Cartridge Based Sample Prep Kit
®
®
Trueprep
AUTO
v2
Plug-in connector