
The device will automatically eject the cartridge holder. Take out the cartridge and place it on the
cartridge stand. Pierce the seal of elute chamber and aspirate out the entire elute into Elute collection
tube (ECT) using Disposable transfer pipette provided in the cartridge pack.
7.4.5 Discard the cartridge as per the instructions in Section 8.3.
7.4.6 Tray replacement
Cartridge holder contains one middle
tray as shown in Figure 13. Steps for replacement of
Replaceable
Replaceable tray are as follows:
Step 1: Removal of the tray:
Remove the
tray from cartridge holder as shown in the Figure 14.
Replaceable
Step 2: Placement of the tray:
2a. Place the
tray in cartridge holder as shown in Figure 15.
Replaceable
i. Inspect the
tray in the cartridge holder
Replaceable
ii. If there are liquids spilled in the tray, take out the tray and discard as per the instructions in
Section 8.3.
iii. Clean the cartridge holder with 70% Isopropyl Alcohol (IPA) as described in Section 8.3.
iv. Replace the removed tray with a new tray and close the cartridge holder (25 Replaceable
®
trays are provided with the
Trueprep AUTO v2
device. More trays, if required, can be
ordered from Molbio).
3 Replaceable
Figure 1 :
tray in cartridge holder
Figure 14: Removal of Replaceable Tray from cartridge holder
13
Replaceable Tray
®
®
Trueprep
AUTO
v2