![Mohr CT100 Скачать руководство пользователя страница 46](http://html.mh-extra.com/html/mohr/ct100/ct100_operators-manual_1823447046.webp)
Operating Instructions
Common types of cable faults
Open faults appear as a pulse upward in the trace (Figure 9):
Short faults appear as a pulse downward (Figure 10):
Mohr CT100 / CT100HF Operator's Manual
37
Figure 9: An open cable fault. The step pulse is completely
reflected by the open termination of the cable.
Figure 10: A short cable fault.