![MOGlabs FSC Скачать руководство пользователя страница 13](http://html.mh-extra.com/html/moglabs/fsc/fsc_manual_1823377013.webp)
2.2 Rear panel controls and connections
9
2.2
Rear panel controls and connections
Serial:
TRIG
FAST OUT
SLOW OUT
MOD IN
POWER B
POWER A
MONITOR 1
MONITOR 2
SWEEP IN
GAIN IN
B IN
A IN
LOCK IN
All connectors are
SMA
, except as noted. All inputs are over-voltage
protected to
±
15 V.
IEC power in
The unit should be preset to the appropriate voltage for your country.
Please see appendix D for instructions on changing the power supply
voltage if needed.
A IN, B IN
Error signal inputs for channels
A
and
B
, typically photodetectors.
High impedance, nominal range
±
2
.
5 V. Channel
B
is unused unless
the
CHB
switch on the front-panel is set to
PD
.
POWER A, B
Low-noise DC power for photodetectors;
±
12 V, 125 mA, supplied
through an M8 connector (TE Connectivity part number
2-2172067-2
,
Digikey
A121939-ND
, 3-way male). Compatible with
MOGL
abs
PDA
and Thorlabs photodetectors. To be used with standard M8 cables,
for example Digikey
277-4264-ND
. Ensure that photodetectors are
switched off when being connected to the power supplies to prevent
their outputs railing.
GAIN IN
Voltage-controlled proportional gain of fast servo,
±
1
V
, correspond-
ing to the full-range of the front-panel knob. Replaces front-panel
FAST GAIN
control when
DIP1
is enabled.
SWEEP IN
External ramp input allows for arbitrary frequency scanning, 0 to
2.5 V. Signal must cross 1.25 V, which defines the centre of the sweep
and the approximate lock point.