Mobilex A/S
Grønlandsvej 5
DK-8660 Skanderborg
www.mobilex.dk
System
Specification
Classification
Class II
Applied Part
Mattress Pad
Input Rating
230V 50Hz, < 10W, 1A
Fuse Rating
250V, F1AL
Air Output
6-7 L/min
Environment Requirements
●
Temperature: Operation 0°C to 40°C
●
Humidity: Operation 30%~75% non-condensing
Safety Standards
CE, Protection Against Class II Electric Shocks
Product label
The product label is located in the bottom of the frame
Sample of the product label only (not original)
5. Cleaning & Maintenance
Cleaning
In this section, we describe the procedures to clean and decontaminate the pump and bubble pad.
It is important to follow these procedures before using the system on patients.
The cleaning task is required at least once a week to maintain personal hygiene.
Pump
●
DO NOT immerse or soak the pump.
●
Check for external damage and move the pump to a cleaning area.
●
Place the pump on a work surface and spray or wipe the outside of the case with quaternary
ammonium solution.
●
DO NOT spray any cleaning solution directly on the surface of the pump.
●
DO NOT use a Hypocarbonate or Phenolic based cleaning solution as this may cause damage
to the case. Allow the solution to incubate for 10 minutes or accordingly as stated by the
cleaning product used.
●
Spray cloth with cleaning solution and clean faceplate. DO NOT allow excess cleaning
solution on faceplate or control panel. (If solution gets inside, damage will occur.)
●
Allow surface to thoroughly dry after cleaning.
●
After the pump is thoroughly cleaned and dried, proceed to plug in the pump and test to see
if it runs normally.
Bubble Pad
●
Any obvious blood spots should be wet thoroughly with 1:9 Hypochiorite solution
●
(1 part bleach to 9 parts water) and allow drying for at least 10 minutes. Then blot with a
clean, damp cloth.
●
Brush or wipe down all surfaces with soap and water before applying any liquid.
●
Allow it to thoroughly air dry. Once the inside is dry, turn it back: wipe down the outside of
the bag with disinfectant.
●
Dry the bubble pad on a SUNLESS area after cleaning.
Rev. 12/15
Page 5 of 6
UM-270016-GB
MOBILEX A/S
Grønlandsvej 5
DK-8660 Skanderborg
Tel: +45 87 93 22 20
www.mobilex.dk
Alternating Pressure
System
Anti-bedsore Anti-decubitus
Mattress with adjustable pump
Pump type P05
Mattress type B01 PVC
Max. user weight 130 kg
Produced