Make sure the inflation tubes are not kinked or tucked under the pad.
●
Before inserting the plug into the outlet, make sure the voltage is compatible.
●
Plug the power cord into an electrical outlet with grounded AC power. If the power cord has
a grounding wire with a grounding pin, then this three-wire plug must be plugged properly
into an outlet and grounded as shown in the figure in the “Grounding description” section.
●
Make sure this product is well grounded.
●
Make sure to disconnect the pump by unplugging the power cord when it is not in use.
Grounding description
This product should be grounded when it is equipped with a cord that contains a grounding wire
with a grounding plug. In the event of an electrical short circuit, grounding reduces the risk of
electric shock by providing an escape wire for the electric current.
Improper use of the grounding plug can result in a risk of electric shock.
Check with a qualified electrician or serviceman if the grounding instructions are not completely
understood, or if in doubt as to whether the product is properly grounded.
If it is necessary to use an extension cord, use only a 3-wire extension cord that has a three-blade
grounding plug and a 3-slot receptacle that will accept the plug on the product. Replace or repair if
damaged.
Function description
●
Turn on the power by pressing the power switch at the right side of the pump. The indicator
of the power switch will come on. The pump starts to pump air into the mattress pad.
●
For time-saving in the first inflation, turn the pressure knob to the full Firm / Hard position. It
requires approximately 10 minutes for a bubble pad to be fully inflated.
●
When mattress pad is fully inflated turn the pressure-adjust knob to set a comfortable
pressure level from soft to firm according to patient’s weight and comfort.
●
It is recommended that the pressure-adjust knob being turned to Firm when the bubble pad
is firstly inflated. Users can then easily adjust the pressure level of the bubble pad to a
desired firmness by themselves or according to the suggestion from a health care
professional.
●
If the pressure is consistently low, check for any leakage occurring in inflation tubes. Contact
your local dealer for replacement of any damaged tubes.
●
The deflating/inflating cycle runs automatically, when pump is turned on.
Transport recommendations
The device was not designed for frequent transportation. If it is necessary to transport it, we
recommend to disassemble it in the way opposite to the assembly description from above. Device
should packed in the way that it is stable and there is no risk of sudden and dangerous movements
of the transported device.
Accessories
There are no accessories provided for this device
4. Technical data
Pump P05
Specification
Dimensions
27,5 x 10 x 10 cm
Weight
1.1 kg
Pressure Range
70 ~ 130 mmHg
Cycle Time
12 min
Mattress Pad
Description
Dimensions
Material
Min./Max. Load
801
Bubble pad w/positioning extension
200×90×6 cm
PVC 0.3 mm
20 kg /130 kg
Rev. 12/15