Connect the male part of the quick connection to the female
part where it is located on the pump.
•
Next, check the air supply hoses so that they are not beneath
the mattress or that they are not folded to a degree of air
pass difficulty.
•
After the final check, connect the power supply to the wall
power supply.
•
Turn on the device.
•
We wait about 12 minutes for the airbag to reach its
maximum air capacity. Once this is done, the Normal
Pressure light will turn green. The Low Pressure lamp will
blink for the entire duration of this inflation.
NOTICE
If Cardio-Pulmonary Resuscitation (CPR) is required in the patient, immediately pull the valve
(CPR) to release air directly from the mattress.
The valve (CPR) is located on the right side at the top of the mattress.
Also remove the connection from the pump for an even faster deflation of the mattress.
13
SPECIFICATIONS
PU
MP
Operating Voltage (V) -Frequency (Hz)
220 VAC/ 50HZ
Air outlet
≥
7 liters /min
Pressure range
30mmHg – 60mmHg
Rotation cycle timing
12 min
Power
≤ 7 W ± 10%
Noise
≤ 30 dB ± 10%
Dimension (ΜxΠxΥ)
27,5 x 13,5 x 9 cm
Pump weight
2 Kg
Maximum inflation time
Approximately 12 minutes
User interface type
All Keys are Touch, only the On / Off button is soft touch
MATT
R
ES
S
Max patient weight
155 Kg
Color
Blue
Material
Nylon TPU & Stretchable PU (cover)
Air Tube Height (inflated)
20 cm
Mattress Height
0,22 mm
Dimensions
200 x 90 cm (18 cells)
Height of foam mattress
8 cm
Connection tube length
150cm
Weight
10 Kg
Содержание 0806288
Страница 2: ...15 Stage to II EL 12 2 ON OFF...
Страница 3: ...3 30 180 kg 10kg 4 L 5 LED 30 2 1 OFF 4 ALARM 5 5 3 0806289 5 3 0806290 DIGITAL 3...
Страница 4: ...1 2 3 4 5 6 CPR 3 4 5 6 7 1 2 4 1 2 5 3 4 6...
Страница 6: ...6...
Страница 7: ...7...
Страница 8: ...1 2 3 8 93 42 8 130648 2198 02 10 2009 T 1 1 93 42...
Страница 9: ...To 2 6 S N LOT 96 98 30 28210 63222 www mobiakcare gr 9...