background image

HEALTH CHECK Ear and Forehead Thermometer

Model No.: 70125 DDX

Instruction Manual
Please read this instruction manual completely before operating.

Package Contents

Quantity

1pc
2pc
1pc

Parts

HEALTH CHECK Thermometer
1.5V (AAA) batteries
User Manual

Thermometer Specification

Power Supply: 

Body Measuring Range:

Ear and Forehead Measure Accuracy:

Object Measuring Range:

Object Measure Accuracy:

Dimensions:

Weight:

Forehead operating Condition:

Ear and Object operating Condition:

Atmosphere pressure:

Service life:

Service life of the battery:

DC 3V (2 x AAA) 
89.6°F 

109.3°F (32.0°C 

42.9°C)

Within range of 95.0°F 

107.6°F (35.0°C-

42.0°C): 
Accuracy = ±0.4°F (±0.2°C)  
Outside of stated range: 
Accuracy = ±0.5°F (±0.3°C)
41.0°F 

139.8°F (5.0°C - 59.9°C)

±2.0°F (1.0°C)
150.0(L)x40.5(W)x 50.0(H)mm(±2mm)
61.0g (without batteries)
59.0°F 

104.0°F (15.0°C 

40.0°C)

With a relative humidity of 20% 

85%

50.0°F 

104.0°F (10.0°C 

40.0°C)

With a relative humidity of 20% 

85%

700hPa 

1060hPa

2 years
6 months based on the use frequency at 
5 times/day

Product Specification

1. Forehead Temperature Symbol
2. Ear Temperature Symbol
3. Object Temperature Symbol
4. Low Battery Symbol
5. Temperature Display
6. Celsius/Fahrenheit Symbol
7. Health Check
8. Cleaning Symbol
9. Memory Symbol
10. Temperature Trending Indication

Note: When the device is powered on all icons are temporarily displayed on 
screen, this is not an indicator of battery power. Battery indicator will display 
constantly when battery level is low.

1. Infrared Sensor
2. LCD Display
3. Power & Scan Button
4. Probe Cap
5. MEM/SET Button
6. Temperature Mode Selection
7. Battery Cover

Installing/Replacing Battery

Remove the battery cover by sliding it toward the bottom of the thermometer.
Insert or replace 2 x 1.5 AAA batteries into the battery compartment, ensuring to 
match the indicating polarity symbols (always use new batteries).
Replace the battery cover on to the thermometer ensuring it snaps closed. 

Note: ‘LO’ display on screen is 

not

an indicator of low power. ‘LO’ is displayed 

when temperature reading is low, this will happen if the device measures the air 
temperature when powering on the device.

Temperature Indicators

Temperature Trend Indication Arrows:

This indicator helps to easily identify 

whether the temperature measurement result is higher or lower than the previous 
temperature taken. If the temperature rises, the indicators will light upwards in 
sequence; if the temperature drops the indicator will light downwards in sequence, 
as shown below. 

Temperature LED Indication:

There are two color LED temperature indicators in this 

unit. When a temperature is taken the corresponding light will illuminate around 

the ‘POWER/SCAN’ button.

Green LED: 

34.0°C (93.2°F) - 37.9°C (100.3°F)

Red LED: 

38.0°C (100.4°F) - 43.0°C (109.4°F)

Celsius and Fahrenheit Setting

When thermometer is turned on press and hold the ‘MEM/SET’ button until you 

hear a beep, to switch between Celsius and Fahrenheit.

Measurement Preparation For Best Results

For more accurate measurements always store the device in same environment it is 
being used, for at least 30 minutes prior. An unexpected fluctuation of ambient 
temperature is likely to lower measuring accuracy. 
External factors may influence ear and forehead temperatures, particularly when 
the individual has been: 

Lying on the ear to be measured or has it covered.

Exposed to very hot or cold temperatures.

Swimming or bathing causing the ear canal to be wet. 

Recently exercising, bathing or other physical exertion as this increase’s 

temperature.

Do not talk or move body during measurement.
Make sure to dry forehead or keep ear canal clean. Unclean ear canal or excessive 
hair may cause inaccurate reading.
The infrared sensor must be kept clean, dry and undamaged to ensure accurate 
results. Please wait 5-15 minutes before taking measurement again.

Replace battery when low battery indicator is “ON”.

Use of this device is not intended as a substitution for consultation with your 
physician. Measurement results are for reference only. Contact your physician if you 
have or suspect and health concerns.
For individuals using hearing aids or ear plugs remove any such device and wait 30 
minutes before taking an ear temperature.

Thermometer Operating Guide

Ear temperature measurement:

Press the ‘SCAN/POWER’ button to turn the thermometer ‘On’. 

To measure temperature via ear, remove the forehead probe cover from the 
unit.

Press the ‘MODE’ button to switch to ear temperature mode, “   ” symbol will 

be displayed on screen. 

Insert the ear probe tip slightly into the ear opening, then press & release the 

‘SCAN’ button to measure the body temperature.

Once you hear a short beep the body temperature will be displayed. Please 

allow 6 seconds before the next reading. The “    ” icon will stop flashing in the 

display.

Forehead temperature measurement:

To measure temperature via the forehead, ensure forehead probe cover is 
attached.

Press the ‘MODE’ button to switch to forehead temperature mode “     ”

symbol will be displayed on screen.

Gently touch the entire tip of the probe to the center of the forehead, 
midway between the eyebrows and the hairline. Ensure no hair is obstructing 

the unit, then press the ‘SCAN’ button, you will hear a short beep and the 

temperature will be displayed.

Note: Results shown from 89.6°F to 109.4°F (32.0°C to 43°

C). ‘HI’ is displayed for 

all results above 109.4°F (43°

C), and ‘LO’ displayed for result below 89.6

°F 

(32.0°C).

Object temperature measurement:

Remove the forehead cap. 

Press the ‘MODE’ button to switch to ‘object’ mode.  “   ” will show up on the 

display. 

Hold the thermometer approximately 1cm away from the object surface, 

press and release the ‘SCAN’ button aiming the measurement sensor at the 

object. 

Once you hear the short beep the object temperature will be displayed. 

Note: If display shows HI/LOW please make sure thermometer is in ‘object’ 

mode.

Care and Maintenance

Lens / Measurement Sensor:  

Gently clean with an alcohol swab. Do not use 

water to wash the thermometer lens directly.

Thermometer: 

Clean with a soft, dry cloth. Do not use water to rinse the device. 

Note: The thermometer is programmed to remind the user to clean the device,    

‘     ’ will flash on screen. Press any button to exit this notification, and please 

clean the unit.

Отзывы: