19.2. External Supply
The VM30-H can be supplied externally via a socket type
Binder
711 (5 pins, female). The
supply voltage range is 8 to 12 VDC with a current consumption below 200 mA. The voltage
source must be able to supply a peak current of 1 A in the moment after switching on the
VM30-H. The power supply input is protected against false polarization.
Operation with external supply is indicated as “Ext.” instead of the battery indicator.
Charging the internal batteries by an external supply voltage is not possible.
Figure 31: Pin assignment of the power supply socket (view from outside)
A protective cover makes the socket waterproof (IP65) when not in use.
20. Loading Default Settings
The instrument can be reset to the initial factory settings by choosing the menu item “6/8: Load
defaults”.
This effects:
•
Reset of the transducer sensitivities to 10.00 mV/ms
-2
•
Reload of the factory calibration values
•
Overwriting the user-defined setups with their default values (see section 9.4)
•
Reset of time and date
•
Reset of other variables like baud rate, beep sound, measuring range, shut-off timer and
battery type
•
Erasing the data memory
21. Hardware Reset
If the VM30-H should not start after longer storage without batteries, it may be necessary to
reset the software. This can be done by pressing the RESET button which is located in the
battery compartment (see Figure 3). After this it may be necessary to load the default settings as
described in section 20. Please check also time and date.