![MMF VM24 Скачать руководство пользователя страница 8](http://html1.mh-extra.com/html/mmf/vm24/vm24_instruction-manual_1813940008.webp)
Besides the frequency range of 10 to 1000 Hz, the VM24 has a further frequency
range for vibration severity measurements from 2 to 1000 Hz. This is suitable for
measurements on slow running machines with a nominal speed of less than
120 min
-1
and on reciprocating engines according to ISO 10816-6.
The corresponding VM24 frequency response graph can be viewed in 4.
Vibration velocity is formed through integration of the acceleration measured by the sen-
sor. The following formula applies to sinusoidal signals:
v
=
a
2
Π
f
It is evident that in the case of constant acceleration (a) the value of velocity (v) drops at
higher frequencies (f). As a consequence the measuring range limits of the velocity mea-
surements are frequency dependent. (5).
The maximum velocity measurable with the VM24 is around 1000 mm/s (RMS). It
yields a frequency independent modulation range of up to 40 Hz. At higher frequen-
cies (RMS) the measuring range limits for acceleration of around 240 m/s² take ef-
6
Figure 4: Frequency response curves for velocity
Figure 5: Amplitude range for velocity (RMS)
1
10
100
1000
10
100
1000
10000
v: 2Hz – 1kHz
v: 10Hz – 1kHz
Integrator
Hz
m
m
/s
1
10
100
1000
10000
0,001
0,010
0,100
1,000
10,000
v: 2Hz – 1kHz
v: 10Hz – 1kHz
Hz
m
m
/s
Содержание VM24
Страница 4: ...2 Figure 1 VM24 with Sensor...