If a Firmware version with a higher number is available on the website, proceed as
follows:
1. Download the firmware file vm2x.hex from the above named internet address.
2. Also download the program “Firmware Updater” from the above named inter-
net address and install it on your PC.
3. Connect the VM24 to the PC using the USB cable and switch it on so that Win-
dows detects it as USB device.
4. Start the “Firmware Updater”, then select the instrument type “VM2x” and the
virtual COM port assigned by the PC. If you are not sure which of the available
COM ports is correct, you can check in the Windows system control manager
located within the instrument manager.
5. Click on “Load” in the “Firmware Updater” and enter the path to the file where
the downloaded firmware file vm2x.hex is located.
6. Within the VM24 “Instrument Settings” select the option “Update Firmware”
and confirm the warning and subsequent hint messages by pressing OK. By car-
rying out this step the old firmware is deleted. The VM24 will then indicate that
it awaits new firmware data from the USB port (36).
28
Figure 35: Firmware Updater
Figure 34: Firmware version
Figure 36: Firmware update