4. BATTERY OPERATION
When the battery is low, the indicator will light up. Change the battery or use an AC adapter
after it turns on.
When battery indication show , charge the battery immediately.
The battery life is approximately 40 hours.
To charge the battery simply plug it into the mains power
.
The balance does not need to be
turned on.
The battery should be charged for 12 hours for full capacity.
The battery indicator flickers when it is charging. When the battery is full, the light stops flickering.
Over time, the battery may fail to hold a full charge. If the battery life becomes unacceptable then
contact your distributor.
5.1 PARAMETER SETTING
5.1 Enter parameter setting
Press
MODE
key during self-checking to enter
parameter setting
mode, after enter setting mode,
press
MODE
key to change setting, press
ZERO
key to confirm setting. Press
key to exit
mode
5.2 Setting the weighing unit
After entering
parameter setting
, the display will show “
f1 unt
”. Press
ZERO
key to enter setting.
Press
ZERO
key to set the weighing unit
.
To return back to
F1 UNT
, press
key to exit
mode.
5.3 Setting the backlight
After entering parameter setting, the display will show “
f1 unt
”. Press
MODE
key until display
show “
F2 El
”. Press
ZERO
key to enter setting. Press
MODE
key to select backlight mode, then
press
ZERO
key to confirm selection. To exit mode, press the
key.
AU
: auto backlight, when load weight is on the pan, backlight will turn on
ON
: always turn on
OFF
: always turn off
Note: when battery is low, the backlight will not function
5. 4 Setting the communication mode
After entering parameter setting, display shows “
f1 unt
”, press
MODE
key until display show “
F3
com
”. Press
ZERO
key to enter setting. Press
MODE
key to select communication mode, press
ZERO
key to confirm selection. To exit mode, press the
key.
6. CALIBRATION
To enter calibration mode, press the
MODE
key while it is self-checking. The display will
show the first function, "
F1 UNT
", press the
MODE
key until display show
“
TECH
”, press
ZERO
key to enter, display will show “
PIN
”, press
SMPL
,
,
TARE
key, and press
ZERO
key to enter technical parameters setting mode.
Press MODE key until display show
“
P2 CAL
”, press
ZERO
, the display will show "
unLoAd
".
Remove all weight from the pan and then press the
ZERO
key to set the initial zero point.
Then put the calibration weigh on pan, after stable, calibration finished.