MM MPB SERIES Скачать руководство пользователя страница 5

Note:

 

the scale may jump by large numbers unexpectedly if small weights are used to set 

the 100% level. For example if only 23.5g is on a scale with 0.5g increments and the scale is 
set to 100%, the display will show 100.00%, however a small change of weight will cause the 
display to jump to 102.13% as one scale division (0.5g) increase to 24.0g will be equivalent 
to a 2.13% increase.

 

Pressing the 

%

 key will return the scale to weighing. 

3.5 Parts counting 

Tare the weight of any container that will be used, leaving the empty container on the scale. 
Place the number of samples on the scale. The number should match the options for parts 
counting, 10, 20, 50, 100 or 200 pieces. When the scale is showing weight, pressing the 

SMPL

 key will start the parts counting function.  

Press the 

SMPL 

key to begin. The scale will show "

SP 10

" asking for a sample size of 10 

parts. You can press 

MODE

 key to select sample quantity: 10, 20, 50, 100, 200 and back to 

10.  

Press the 

ZERO

 key when the number shown matches the number of parts used for the 

sample. As more weight is added the display will show the number of parts (pcs).  

Press the 

MODE

 key to display unit weight (g/pcs), total weight (kg) or the count (pcs). 

Press the 

SMPL

 key to exit parts counting mode. 

3.6 Accumulation 

The weight displayed will be stored in memory when the 

PRINT

 key is pressed and the 

weight is stable, the display will show "

ACC 1

", then the total in memory for 2 seconds 

before returning to normal. If the RS-232 interface is installed, the weight will be sent to a 
printer or PC. 

Remove the weight, allowing the scale to return to zero and put a second weight on. Press 

PRINT

, the display will show 

ACC 2

 

and then show the new total. 

To view the totals in memory, press the 

PRINT

 key when the scale shows zero. The display 

will show the total number of items "

ACC xx

" and the total weight before returning to zero. 

The totals will also be printed via the RS-232 interface. 

To erase the memory press 

PRINT

 to view the totals, followed by the 

SMPL 

key to clear the 

memory. 

3.7 Calibration 

In normal weighing mode, empty the pan, hold the 

MODE 

key for 3 second, display will 

show 

unload

”. Wait for several seconds until the display shows “

LOAD

”. Put the calibrate 

weight on the pan. After 

STABLE 

indicator turns on, the window will show “

pass

”.

 

After self-

checking

the calibration will be complete.

 

Note: you can use half capacity or full capacity calibrate weight to calibrate the balance. You 
don

’t need key in the weight value, as the balance will automatically detect the weight. 

-3- 

 

 