MKS Instruments UK Ltd 

V2000-C/E Hardware Manual – SP102005.102 March 2013

 

 

 

26 

 

 

CAUTION

 

Check the complete installation thoroughly before proceeding. 

 

 

Start up 

Set the Remote Vacuum Controller (RVC) “Interlocks” key switch to ON and the power switch to On (I) 
 
On your PC, start Process Eye Professional. The software will detect that the RVC2 is fitted and the Remote 
Vacuum Controller status and control panel will be available at the bottom of the Process Eye Professional 
window.   
 
To start the Vision 2000C/E vacuum system, click on the turbopump button.   
The colour of the button indicates the status:  

Red = off 
Yellow = starting 
Green = running at full speed 

 
More information about the RVC operating software can be found in the manuals for the RVC2 and Process Eye 
Professional.  
 
The Vision 2000C/E may also be started from the RVC2 directly. This is useful if the PC is disconnected or the 
Process Eye software is not running. To start the system from the RVC2, briefly press the PUMPS switch on the 
front panel.  Note that this control can also be used to turn the turbopump OFF. 
 
As soon as the Vision 2000C/E is started the SurgeProtector foreline valve should open (if tool vacuum is <3torr) 
and the turbopump will start. 
 
The turbo pump controller will indicate when the pump is up to speed and the Turbopump indicator in the RVC 
window will change from yellow to green.  Note that the turbopump controller will be programmed to a set-point 
of 95% of rated speed. 
 
After the turbopump has reached full speed, you may switch the RGA filament on, but we recommend waiting at 
least 10 minutes after the adsorbed gas load in the RGA vacuum chamber to reduce before switching a filament 
on. 
 
Switch the RGA filament on and look at the RGA chamber background spectrum using a Process Eye 
Professional Bar-Chart recipe (refer to the Process Eye Professional documentation for further details). 
 
The pressure is likely to be quite high but should be falling. If the pressure is very high switch the filament off and 
wait a little while for the pressure to fall.  
 
Now, check that the process valves work. Ensure that there is a good vacuum in the process chamber. The 
pressure in the process chamber must be below the corresponding sample-valve pressure set-point to permit 
you to open a sampling valve. 
 
With the Remote Vacuum Controller in the manual mode and a filament on, open the process valve. The process 
valve will open and you will be able to see a spectrum of the process chamber. 
 
Once you have established that the process valve is working correctly, close the valve.  Minimizing exposure to 
corrosive process gases when not actively monitoring them will protect RGA system operating life. 
 
 

Содержание Vision 2000-C

Страница 1: ...Vision 2000 C E Manual SP102005 102 March 2013...

Страница 2: ...ticularly relating to any omissions or inaccuracies you may discover You can send your comments to MKS Instruments Spectra Products Cowley Way Weston Road Crewe Cheshire CW1 6AG United Kingdom 44 0 12...

Страница 3: ...MKS Instruments UK Ltd V2000 C E Hardware Manual SP102005 102 March 2013 3...

Страница 4: ...MKS Instruments UK Ltd V2000 C E Hardware Manual SP102005 102 March 2013 4...

Страница 5: ...MC and safety issues The vacuum system on which the analyser RF head is mounted must be earthed to a protective earth preferably to the same protective earth as the control unit Operation The equipmen...

Страница 6: ......

Страница 7: ...ecting the foreline and SurgeProtector valve 19 Heater jackets 20 MicroVision2 installation 20 ELECTRICAL INSTALLATION 21 Equipment rack 22 Interconnecting cables 22 CONNECTING TO THE PROCESS TOOL 23...

Страница 8: ...these boxes is described below WARNING WARNING boxes are used where failure to observe the instructions could result in personal injury or death CAUTION CAUTION boxes are used where failure to observ...

Страница 9: ...ar panel of the Microvision2 refers to Read all instructions carefully before use The control and signal ports are designed for connection to MKS Instruments accessories via MKS Instruments cables The...

Страница 10: ...mm H R C ceramic 250V AC characteristic T and compliant with IEC 127 Details of fuse types and ratings can be found in the RVC2 manual and printed on the rear panel Electrical Connections The Vision 2...

Страница 11: ...om the tool foreline when the RGA system pump is off and when the tool foreline pressure is above 3 torr WARNING The RGA system must never be operated without exhausting to a hazardous gas scrubbing s...

Страница 12: ...uments UK Ltd V2000 C E Hardware Manual SP102005 102 March 2013 12 System Options Fig 1 3 Valve Etch Configuration Fig 2 4 Valve Fast Response LPCVD Configuration Fig 3 4 Valve High Pressure CVD Confi...

Страница 13: ...r local MKS Instruments facility or sales service representative Carefully unpack the various parts of your Vision 2000 C E system Again check for any signs of damage Find the shipment report and chec...

Страница 14: ...n 2000C E has makes little difference to its installation On the face of the UniBloc inlet valve that connects to the RGA chamber is a ceramic CIS closed ion source coupling Exchanging an inlet valve...

Страница 15: ...MKS Instruments UK Ltd V2000 C E Hardware Manual SP102005 102 March 2013 15 Dimensions 490mm 225mm 265mm 340mm 166mm...

Страница 16: ...depending on working pressure 722 Baratron Gauge 2 LoVacOK 890B Gauge 3 UniBloc inlet assembly 4 Purge regulator assembly 3 5 to 8Bar Argon or dry Nitrogen supply required 5 System Interface Module 6...

Страница 17: ...e Manual SP102005 102 March 2013 17 Item Ref Description 8 Pneumatic Manifold assembly 60 80 psig dry air 9 Fore line connecting flange 10 Cable loom supporting bracket 8 9 10 Pneumatics supply Use ad...

Страница 18: ...MKS Instruments UK Ltd V2000 C E Hardware Manual SP102005 102 March 2013 18 Item Ref Description 15 V1 16 V2 17 V3 18 V4 19 UniBloc mounting face 20 16 15 17 18 19...

Страница 19: ...g response time can be improved by using a length of 1 8 OD PTFE tubing from the point to be sampled to the face of the UniBloc Be sure that the tubing material is compatible with the process gases to...

Страница 20: ...must be strictly followed Heater jackets The Vision 2000C E is supplied with two heater jackets fitted one on the main RGA chamber and another around the inlet valve assembly There is no need to remov...

Страница 21: ...s and regulations The electrical installation should be carried out after the vacuum system installation Please follow the next sections in sequence Electrical specification See the RVC2 and Microvisi...

Страница 22: ...n protective conduit e g plastic trunking that provides a good degree of protection according to the location of installation All cables are labelled at both ends All the connectors have been chosen t...

Страница 23: ...100 120 Volts AC 50 60 Hz 15 Amps 220 240 Volts AC 50 60 Hz 13 Amps LINE BROWN NEUTRAL BLUE EARTH GREEN YELLOW Computer connection Vision 2000C E systems are used in conjunction with MKS Instruments...

Страница 24: ......

Страница 25: ...mputer software The operation of the vacuum system controller is fully described in the Remote Vacuum Controller manual The pumps are switched on from the Remote Vacuum Controller window or from the R...

Страница 26: ...r will indicate when the pump is up to speed and the Turbopump indicator in the RVC window will change from yellow to green Note that the turbopump controller will be programmed to a set point of 95 o...

Страница 27: ...unt recipe in Process Eye and check that the probe gas is set to mass 4 for helium or the appropriate mass if you are using another gas As the Vision 2000C E uses a Microvision2 there can be an audio...

Страница 28: ...CAUTION The Electron Multiplier SEM MUST NOT be operated at temperatures above 90 o C Do NOT bake with the SEM selected After baking ends wait at least 2 hours before using the SEM The electron multip...

Страница 29: ...ween the two sequential orifice stages O2 1 and O2 2 The inlet pressures that can be sampled properly are determined by this configuration and the size of the orifices Again differential pumping is ob...

Страница 30: ......

Страница 31: ...the two flying leads coming from the gauge connector and measure the voltage Ensure that the process chamber pressure is below 1milli torr then check the meter reading If the indicated voltage is gre...

Страница 32: ...oval For system with compression springs note the following tightening procedure tighten these screws until the top of the screw head is flush with the top surface of the ceramic coupling springs will...

Страница 33: ...tron multiplier SEM detector will also require replacement In extreme cases this can be as soon as 3 months but 1 2 years is more typical For non corrosive gas sampling SEM performance is often accept...

Страница 34: ...f 6 o clock the standard analyser alignment pip orientation is at 9 o clock Remove the six x 28 bolts Carefully withdraw the analyser from the vacuum chamber Leave the old copper gasket in place until...

Страница 35: ...rned delays in providing this completed form will lead to delays in the servicing of the instrument Securely package all items to be returned using the original packaging where possible and send to th...

Страница 36: ...ease ensure that we have this information before we receive the equipment A copy should also be given to the carrier FAILURE TO COMPLETE THIS FORM OR COMPLY WITH THE PROCEDURE WILL LEAD TO DELAYS IN S...

Страница 37: ...ese substances Action to be taken in the event of human contact or spillage I hereby confirm that the only toxic or hazardous substances that the equipment specified above has been in contact with are...

Отзывы: