Chapter Three: Overview
Rear Panel Connections and Controls
23
Chapter Three: Overview
Rear Panel Connections and Controls
Pressure Sensor Connector
These connectors are used to interface the PDR with the pressure gauge(s). They provide power
to gauges requiring it, and receive the pressure signal back from the gauge. Refer to Table 9,
page 28, for the connections.
Decimal Point Selector
These are 4PST rocker switches used to set the display decimal point to the appropriate position for
the pressure range being measured. The following chart shows switch settings for various full scale
pressure values.
Decimal Point Selector Switch
Gauge F.S.
Switch Selection
1
2
3
4
1 MM
ON
OFF
OFF
OFF
10 MM
OFF
ON
OFF
OFF
100 MM
OFF
OFF
ON
OFF
1000 MM
OFF
OFF
OFF
ON
10000 MM
OFF
OFF
OFF
OFF
Table 7: Decimal Point Selector Switch
Note
Closing more than one switch (per channel) will produce multiple decimal
points and cause interaction between channels.