Chapter Two: Installation
27
Piping Considerations and Port Connections
General
Position a line tap above or to the side of the process line, and position the sensor above
the line tap
This allows any liquids which may be present to drain into the process line.
Install the interconnecting piping so that it slopes
away from
the sensor to the process line
A slope of 1 inch per foot (8 centimeters per meter) is adequate.
Purge the interconnecting piping (close to the process tap) when needed to prevent
sediment buildup
When purging, be sure to isolate the sensor from the interconnecting piping.
Sensor Fittings
The following additional fittings are available for the Type 631D transducer:
½” (12.7 mm) tubing
Swagelok
®
8-VCR
®
female
NW-16-KF
1½” Tri-Clover
®
2” Tri-Clover
Special Note for ½” (12.7 mm) Tubing
The standard tubing for the 631D transducer is straight ½” (12.7 mm) tubing. The Type 631D
transducer has only one pressure port (Px).
To connect the 631D transducer to a Swagelok
®
Ultra-Torr
®
compression fitting:
1. Ensure the tubing is clean and free of axial scratches.
2. Insert the tubing through the compression nut and O-ring all the way to the shoulder.
3.
Tighten the nut.
Содержание Baratron 631D
Страница 3: ...1039281 001 Rev B 3 13 MKS Baratron Type 631D High Temperature Pressure Transducer...
Страница 28: ...Chapter One General Information 20 This page intentionally left blank...
Страница 36: ...Chapter Two Installation 28 This page intentionally left blank...
Страница 44: ...Chapter Three Overview 36 This page intentionally left blank...
Страница 50: ...Chapter Four Operation 42 This page intentionally left blank...
Страница 54: ...Chapter Five Optional Solid State Relay System 46 This page intentionally left blank...
Страница 61: ...Appendix B Model Code Explanation 53...
Страница 62: ...Appendix B Model Code Explanation 54 This page intentionally left blank...