Chapter Two: Installation
24
Environmental Requirements
Caution
DO NOT use a thermal blanket on the 631D unit. The 631D
transducer is designed to maintain its specified operating
temperature without external assistance.
Operating ambient temperature:
150
C units:
10° to 50° C (50° to 122° F)
200
C units:
10° to 50° C (50° to 122° F)
Caution
Airflow of at least 50 ft/min is REQUIRED for 150
C units if the
ambient temperature is 45
to 50
C (113 to 122
F).
Airflow of at least 150 ft/min is REQUIRED for 200
C units if the
ambient temperature is 40
to 50
C (104 to 122
F).
Ventilation requirements include sufficient air circulation
Caution
All high temperature equipment, including the 631D transducer,
requires sufficient air circulation to prevent heat buildup. If the 631D
transducer is placed in an environment with still air or insufficient air
circulation, the temperature of the air surrounding the transducer
will rise. You must provide sufficient air circulation to prevent heat
build up around the 631D transducer.
Storage Humidity Range: 25% to 95% relative humidity, non-condensing
Storage Temperature Range: -20
to +80
C (-4
to 176
F)
Содержание Baratron 631D
Страница 3: ...1039281 001 Rev B 3 13 MKS Baratron Type 631D High Temperature Pressure Transducer...
Страница 28: ...Chapter One General Information 20 This page intentionally left blank...
Страница 36: ...Chapter Two Installation 28 This page intentionally left blank...
Страница 44: ...Chapter Three Overview 36 This page intentionally left blank...
Страница 50: ...Chapter Four Operation 42 This page intentionally left blank...
Страница 54: ...Chapter Five Optional Solid State Relay System 46 This page intentionally left blank...
Страница 61: ...Appendix B Model Code Explanation 53...
Страница 62: ...Appendix B Model Code Explanation 54 This page intentionally left blank...