General
Chapter Four: Operation
36
Suggested Pressures for Reading and Control
Full Scale Range
(Torr)
Lowest Suggested Pressure
for Reading (Torr)
Lowest Suggested Pressure
for Control (Torr)
0.05*
2.5 x 10
-5
2.5 x 10
-4
0.1 **
5 x 10
-5
5 x 10
-4
1
5 x 10
-4
5 x 10
-3
10
5 x 10
-3
5 x 10
-2
100
5 x 10
-2
5 x 10
-1
1K
5 x 10
-1
5
** Only available with Types 627 and 629 transducers
** Only available with Types 627, 628, and 629 transducers
Table 11: Suggested Pressures for Reading and Control
Содержание Baratron 622A
Страница 8: ...List of Figures and Tables vi ...