Chapter Two: Installation
Product Location and Requirements
23
Product Location and Requirements
Types 622 and 626:
•
Ambient operating temperature should remain between 0° C and 50° C (32° F and 122° F)
•
Input required is ±15 VDC (±5%) @ 35 mA (maximum) with <20 mV p-p noise and ripple
Type 627 and 629:
•
Ambient operating temperature should remain between 15° C and 40° C (59° F and 104° F)
•
Input required is ±15 VDC (±5%) @ 250 mA (maximum) with <20 mV p-p noise and ripple
Type 628:
•
Ambient operating temperature should remain between 20° C and 70° C (68° F and 158° F)
•
Input required is ±15 VDC (±5%) @ 500 mA (maximum) with <20 mV p-p noise and ripple
For additional product requirements refer to Appendix A: Product Specifications, page 41.
Содержание Baratron 622A
Страница 8: ...List of Figures and Tables vi ...