Setup
Chapter Two: Installation
26
ZERO Potentiometer
Type 622
Type 626
Type 627, 628, and 629
(Type 629 has a 9-pin connector)
Type 622
Pressure
Transducer
+15 VDC-35 mA
PIN SIGNAL
1 Pwr Com
2 Sig Com
3 Sig Output
4 -15 VDC
5 +15 VDC
Chassis
ZERO
ZERO
Type 62X
Pressure
Transducer
+15 VDC-250 mA
PIN SIGNAL
2 Sig Output
5 Pwr Com
6 -15 VDC
7 +15 VDC
12 Sig Com
15 Chassis
ZERO
Type 626
Pressure
Transducer
+15 VDC-35 mA
PIN SIGNAL
2 Sig Output
5 Pwr Com
6 -15 VDC
7 +15 VDC
12 Sig Com
15 Chassis
1
5
R
BARATRON
R
BARATRON
Figure 4: Location of the Zero Pot on Transducer Endcaps
To properly zero a transducer, follow this procedure:
1. Install the transducer a system with a power supply/readout.
2. Power the transducer and allow it to warm up. Warm up time is:
•
15 minutes for Types 622 and 626 transducers
•
2 - 4 hours for Types 627, 628, and 629 transducers (
≥
1 Torr)
•
8 hours for the 0.05 and 0.1 Torr range units
3. Pump the unit down to a pressure below its resolution (refer to Table 6 for recommended
pressure levels).
4. Using a small screwdriver, adjust the ZERO pot until the readout displays zero (0000).
Highest Pressure Suggested for Proper Zero Adjustment
Full Scale Range (Torr)
Highest Pressure for Zero Adjustment (Torr)
0.05*
< 2.5 x 10
-6
0.1**
< 5 x 10
-6
1
< 5 x 10
-5
2
< 1 x 10
-4
10
< 5 x 10
-4
100
< 5 x 10
-3
1K
< 5 x 10
-2
* Only available with the Type 627 and 629 transducers
** Only available with Types 627, 628, and 629 transducers
Table 6: Highest Pressure Suggested for Proper Zero Adjustment