Chapter Six: Remote Operation
How To Configure the Valve Parameters
85
How To Configure the Valve Parameters
In order for the 651 controller to work properly with your valve, you need to check several valve
parameters. All of the parameters are set to an initial value, so it may be unnecessary for you to
change any entries.
How To Check the Valve Selected
1. To check the type of valve selected, issue the request:
R23
ENTER
The 651 controller responds with the message:
J
type
where
type
=
1 for a standard 253
2 for a fast 253
3 for a 653
How To Change and Calibrate the Valve
1. To change and calibrate the valve, issue the command:
J value
ENTER
where value =
1 for a standard 253
2 for a fast 253
3 for a 653
Caution
During the calibration process, the 651 controller learns the valve by
moving it from the opened to the closed position. Be sure your
system is set up to allow the valve to move from full open to full
closed.
Содержание 651C
Страница 12: ...List of Figures x...
Страница 14: ...List of Tables xii...
Страница 34: ...Customer Support Chapter One General Information 20 This page intentionally left blank...
Страница 74: ...How To Calibrate Span of the A D Converter Chapter Five Local Operation 60 This page intentionally left blank...
Страница 110: ...Digital Logic Control Chapter Six Remote Operation 96 This page intentionally left blank...
Страница 138: ...RS 232 Configuration Appendix D Type 651 Displayless Unit 124 This page intentionally left blank...
Страница 154: ...Request and Response Reference Appendix F Command and Request Reference 140 This page intentionally left blank...
Страница 168: ...Returning the Type 651 655 Unit for Firmware Revision General Information Page 4 of 8 This page intentionally left blank...
Страница 170: ...Safety Information Page 6 of 8 This page intentionally left blank...