Chapter Four: Operation
Additional Procedures for the 0.1 Torr Sensors
43
C. Turn the shaft an extra ½ turn in the same direction.
D. Turn the control back slowly until the same output just appears.
E. Carefully turn the shaft an additional ¹/
8
turn in the same direction.
This will center the pot in the middle of a step for maximum stability.
Note
The above method is the only method that accurately finds the center of a
wire. The technique of turning the zero control shaft clockwise until it
shifts, turning it counterclockwise until it shifts, and positioning it
midway does not work properly.
5. Adjust the
COARSE
and
FINE
zero controls on the 270 instrument to bring the display
(or output) to precisely zero.
When using the Fine control, you may need to down range the 270 instrument to the x0.1
range to make the final setting. Do not use any other range settings.
Using a Low Range Pressure Sensor with the Remote Zero Option
If you are using a pressure sensor less than 1 Torr Full Scale and your 270 instrument is equipped
with the Remote Zero option, you should consult an MKS Service Center. Your 270 instrument
may need to be modified.
Please refer to the back cover of this manual for a list of the authorized MKS Calibration and
Service Centers.