Chapter Three: Overview
Front Panel Controls
27
RESPONSE Toggle Switch
This switch controls the electrical bandwidth of the instrument. The three positions are:
RESPONSE Switch Settings
Setting
Hz
Milliseconds
FAST
165
1
NORMAL
4
40
SLOW
0.4
400
Table 5: RESPONSE Switch Settings
HEATER Lamp
The
HEATER
lamp is illuminated when power is applied to the sensor heater.
HEATER Toggle Switch
This switch controls power to the heaters. The switch has two positions:
REG
and
OFF
.
When the switch is in the
REG
position, power is applied to the heaters in a sensor to maintain
a preset temperature. This minimizes zero and span drift due to ambient temperature variation.
This switch is a locking type switch and the handle must be pulled back to change positions.
When the switch is in the
OFF
position, the heaters are not powered.
POWER Switch
This switch controls power to the 270 instrument.