Labels
Chapter Three: Overview
34
The jumper settings for various system configurations are listed in Table 10.
Jumper Summary
Desired Settings
Jumper Position
J3
J4
J5
0 to 5 VDC, TTL Logic
VIN
0 to 5 V
5 VDC*
0 to 10 VDC, TTL Logic
VIN
0 to 10 V
5 VDC*
0 to 5 VDC, Open Collector
VIN
0 to 5 V
O.C.
0 to 10 VDC, Open Collector
VIN
0 to 10 V
O.C.
4 to 20 mA, TTL Logic
4 - 20
0 to 5 V
5 VDC*
4 to 20 mA, Open Collector
4 - 20
0 to 5 V
O.C.
* Connect a 5 VDC supply to pin 14 on the Interface connector (refer to
Table 6, page 27).
Table 10: Jumper Summary
Labels
Serial Number Label
The serial number label lists the serial number and the product model code. The label also
displays the CE mark signifying compliance with the European CE regulations.
1249A
Serial #:
Model #:
0123456789
MKS Instruments, Inc. Made in the USA
Figure 8: Serial Number Label