OPERATOR MANUAL
EXLRT
Ref: XLR-104
© 2020 MK Test Systems Ltd
Page 21 of 28
11
S
INGLE
S
HOT
M
EASUREMENTS
The icon appears on the LCD screen to indicate that the ExLRT is in single shot mode.
Initiating a single shot test
Refer to the table below on how to initiate a test.
CONNECTED
CABLES
TEST
WAITING
INDICATOR
HOW TO INITIATE A TEST
Loop cable
ONLY
Loop cable
green led
flashing
Press the green button on the In-Line
controller or press the
button to
initiate a test.
When the test stops, the loop value or
the loop error will latch on the screen.
If an error is shown, then press the loop
inline controller to clear the error to
return to flashing green status.
Loop cable
AND
Joint cable
Loop cable
green LED
flashing
Joint probe
green LED
flashing
Press the green button on the In-Line
controller. This will perform a loop only
test (Joint indicator and data is cleared)
When the test stops, the loop value will latch on the screen and
both the loop and joint LED’s will flash green.
If an error is shown, then press the loop inline controller red
button to clear the error to return to flashing green status.
OR
Connect the tips of
the joint probes to
the UUT to initiate
a joint test.
Both loop and joint measurements will be cleared for the duration
of the test
When the test stops, the ExLRT will latch the valid measurements
until the next measurement is started
If an error is reported for a given test, then the ExLRT will latch the
error until the red button is pressed for that measurement type.
During this latched period, the other measurement type will clear
its indicator.