40
6
Operation
6.4 Tilting the tabletop laterally
WARNING
When you tilt the tabletop laterally with a patient on the operating table, make sure to
use the fixture for the accessory of the Mizuho operating table. The patient may fall
from the operating table.
Tilting to the left
1.
Press
first and then
.
The tabletop tilts to the left in the view from the head side.
Left down
Tilting to the right
1.
Press
first and then
.
The tabletop tilts to the right in the view from the head side.
Right down
NOTE
•
The maximum angle
achieved in the left down
and right down position is
25° from the level position.
25°
25°
25°
25°
•
If the tabletop is tilted
laterally to its maximum
extent, the right-left sliding
position may move a little.
NOTE
About level correcting
When the lateral tilt does not
return to the level position
even if the Level switch is
pressed, then press the
right and left Lateral tilt
switches to operate them
to greatly exceed the level
position. The lateral tilt’s
level position will be reset,
and it will recognize the
correct level position. If the
correct level position is not
recognized even though
this operation is performed,
then request repairs from
your distributor or Mizuho.