Safety precaution
CAUTION
●
Moving and transporting
•
Do not move the operating table with a patient on it.
•
Follow the procedures below to move the operating table.
* Before moving the operating table, disinfect the entire operating table in order to
prevent infection.
1. Turn off the power and disconnect the power cord from the medical grade outlet.
2. Check if the handles and levers are in fixed positions, and each section is fixed
firmly.
3. Unlock the brakes, and move the operating table.
* Do not move the operating table by pulling on the head plate or leg plate.
•
The operating table should be transported with the following conditions met.
1. Disinfect the entire operating table before transporting it.
2. Take measures to prevent it from tipping over, such as lowering the tabletop to
the bottom position.
3. Actuate the brake.
4. Suitably position cushioning on the product to prevent it from getting damaged
during transport.
5. Store the product in a container so that it does not get exposed to dust, and the
weather.
●
Disposal
•
In accordance with the European Union Waste Electrical and Electronic Equipment
(WEEE) Directive, all electrical components and batteries must be disposed of in
accordance with local regulations. Please contact your local distributor for proper
disposal.