Instructions for Use 

Trade Name: Rotatable Micro Scissors DT 

 

Warning 

Please read these instructions carefully prior to using this 
product. Product should be used according to these 
instructions and pay close attention to the safety of 
patients. Not doing so may give rise to serious problems 
or adverse events. 

For the US market

 

Do not reuse the device when it is used on a patient with 
or suspected of having Creutzfeldt-Jakob Disease (CJD) 
or variant CJD (vCJD). 

For the market outside the US 

When the device is used on a patient with or suspected of 
having CJD or vCJD, make sure to adhere to the most 
recent and updated restrictions available in each country 
and/or region for its reuse. Be cautious of the possibility of 
secondary infection. Refer to www.a-k-i.org or AAMI 
Standard ST79 for more information related to cleaning 
and sterilization.

 

 

Contraindication / Prohibition

 

1. Use for intended purpose only 

Use tools for their intended purposes only. Incorrect use 

could cause breakage of this product. 

2. Prohibition of use of chemicals 

Avoid exposing this product to chemicals. Doing so could 

damage this product due to corrosion. 

3. Prohibition of secondary processing of this product 

Do not apply any secondary processing to this product, for 

example, do not apply impacts or vibration markings to 
the surface of this product. Doing so could break this 
product. 

4. Handle with care 

Handle this product with care, as it can be deformed or 

damaged. Rough handling could significantly reduce the 
service life of tools and appliances. 

5. Prohibition of use of polishing powder and wire wool 

When cleaning this product, do not attempt to polish its 

surfaces with rough polishing powder or wire wool. This 
could cause scratches on the surfaces of this product and 
result in rust or corrosion. 

6. Prohibition of use of acid and household detergents 

Use only medical neutral detergents (pH 6 to 8) to clean 
this product. Do not use any acid or household detergent. 
Washing this product with an improper detergent could 
result in discoloration or corrosion. 

7. Do not use low temperature hydrogen peroxide gas 

plasma sterilization. 

This product is not suitable for sterilization with hydrogen 
peroxide low temperature gas plasma. It may discolor the 
surface of the product or affect the feature of the product.

 

Shape / Structure 

 
 
 
 
 
 

45 deg Up                        

Straight 

 
 
 
 
 

Curved                            

125 deg Back Cutting 

 
 
 
 
 

Code No. 

Product Description 

07-829-41 

Evans Rotatable Micro Scissors 11cm WL, 45deg. Up,  
DT 

07-829-42 

Evans Rotatable Micro Scissors 15cm WL, 45deg. Up, 
DT 

07-829-43 

Evans Rotatable Micro Scissors 11cm WL, Str. DT 

07-829-44 

Evans Rotatable Micro Scissors 15cm WL, Str. DT 

07-829-45 

Evans Rotatable Micro Scissors 11cm WL, Cvd. DT 

07-829-46 

Evans Rotatable Micro Scissors 15cm WL, Cvd. DT 

07-829-47 

Evans Rotatable Micro Scissors 11cm WL, 125deg. Back 
Cutting, DT 

07-829-48 

Evans Rotatable Micro Scissors 15cm WL, 125deg. Back 
Cutting, DT 

Material: Stainless Steel, Gold Plating 

Intended Purpose 

This product is a surgical instrument used to cut tissue, cloth, 
and a suture. It is composed of two flutes that have 
rotational axes, and it is operated by closing the flutes over 
materials which are supposed to be cut. 

Instructions for use 

Before using this product, inspect, wash, and sterilize in 
accordance with these instructions. 

Cautions / Warnings 

1.

 

Warning 

Device must be sterilized by users in accordance with our 
recommended sterilization procedures or the validated 
sterilization conditions for which validity has been proven by 
medical organizations in each country or region. 

2.

 

Defect/Adverse event 

Defect 

- Corrosion and pitting caused by use of chemicals 
- Damage or breakage caused by the corrosion or pitting  
- Plating adhesion failure 

 Adverse events 

- Broken pieces of metal from the damaged instrument 

falling into the patient. 

3.

 

Important fundamental cautions 

This product has a removable cleaning port. The cleaning 
port is attached to clean the lumen . When cleaning is 
complete, remove it. Refer to

Figure 1

 

 

 

 
 
 
 
 

Be sure to read this instruction manual before using this product. 
Keep this manual available for reference when needed.  

The cleaning port is  
attached to the lumen . 

The cleaning port is  
removed from the lumen. 

Cleaning port 

Figure 1

 