The recommended sterilization parameters are as follows, 

 

Sterilization 

method 

Pre-vacuum steam sterilization  

(Autoclave sterilization) 

Sterilization 

conditions 

Sterilization temp. 

Retention time 

132°C / 269.6°F 

4 min 

134°C / 273.2°F 

3 min 

 

For the US market

 

FDA has not approved or cleared medical devices, 
including sterilizers, for the intended use of reducing the 
infectivity of TSE agents (i.e., prions). 

For the markets outside the US 

When the device is used on a patient with or suspected 
of having CJD or vCJD, make sure to adhere to the most 
recent and updated restrictions available in each country 
and/or region for its reuse. 

Maintenance and check by agents 

For safe use of this instrument, inspect device prior to and 
after each use. Alternative or no review and/or inspection 
may cause injury to the patient and/or healthcare worker 
and may decrease the performance and function of this 
device. Additionally, it is recommended to schedule a 
periodic inspection through your local distributor or the 
manufacturer. 

Packing

 

1 peace per pack 

Matters related to warranty period 

MIZUHO Corporation will repair defective parts of this 
product without charge for one year from the date of 
delivery/installment except for cases of damage caused by a 
third party’s repair, act of nature, improper use or intentional 
damage. All other warranty terms and conditions are subject 
to regulations of MIZUHO Corporation. 

N

ame and address of manufacturer 

 

MIZUHO Corporation 
3-30-13 Hongo, Bunkyo-ku, Tokyo 113-0033, Japan  
http://www.mizuho.co.jp

 

 