The recommended sterilization parameters are as follows,
Sterilization
method
Pre-vacuum steam sterilization
(Autoclave sterilization)
Sterilization
conditions
Sterilization temp.
Retention time
132°C / 269.6°F
4 min
134°C / 273.2°F
3 min
For the US market
FDA has not approved or cleared medical devices,
including sterilizers, for the intended use of reducing the
infectivity of TSE agents (i.e., prions).
For the markets outside the US
When the device is used on a patient with or suspected
of having CJD or vCJD, make sure to adhere to the most
recent and updated restrictions available in each country
and/or region for its reuse.
Maintenance and check by agents
For safe use of this instrument, inspect device prior to and
after each use. Alternative or no review and/or inspection
may cause injury to the patient and/or healthcare worker
and may decrease the performance and function of this
device. Additionally, it is recommended to schedule a
periodic inspection through your local distributor or the
manufacturer.
Packing
1 peace per pack
Matters related to warranty period
MIZUHO Corporation will repair defective parts of this
product without charge for one year from the date of
delivery/installment except for cases of damage caused by a
third party’s repair, act of nature, improper use or intentional
damage. All other warranty terms and conditions are subject
to regulations of MIZUHO Corporation.
N
ame and address of manufacturer
MIZUHO Corporation
3-30-13 Hongo, Bunkyo-ku, Tokyo 113-0033, Japan
http://www.mizuho.co.jp