MIZUHOSI 2009
NW0510 Rev. D
29
•
Call Mizuho OSI Customer Resource Group for service or parts (possible hand
pendant and/or 10 Amp motor-control PCB).
Trendelenburg or reverse trendelenburg not functioning:
•
Press and the hold hand pendant TREN or REV.TREN button. Observe tabletop
movement.
•
Verify the hand pendant is properly plugged into a hand pendant receptacle.
Observe hand pendant buttons illuminate green.
NOTE:
The table power switch must be turned off before disconnecting or
connecting the hand pendant. Once reconnected, turn power on and deploy floor
locks (all four locking feet make firm contact with the floor).
•
With the power switch off, plug the hand pendant into the other hand pendant
receptacle. Turn power switch on. Observe hand pendant buttons illuminate green.
•
Verify power cord is plugged into a properly grounded AC receptacle.
•
If on battery power, verify proper charge; observe battery status LED illuminates
green on the hand pendant.
•
Verify power switch is on.
•
Verify floor locks are deployed (all four locking feet make firm contact with the floor);
observe LOCKED WHEN LIT indicator illuminates green on the control panel and
the FLOOR LOCK ON LED illuminates on the hand pendant.
•
Verify if other hand pendant buttons are functional. If not, replace hand pendant.
•
Call Mizuho OSI Customer Resource Group for service or parts (possible hand
pendant and/or 10 Amp motor-control PCB).
Left lateral roll or right lateral roll not functioning:
•
Press and hold hand pendant LEFT LATERAL ROLL or RIGHT LATERAL ROLL
control button. Observe tabletop movement.
•
Verify hand pendant is properly plugged into a hand pendant receptacle; observe
hand pendant control buttons illuminate green.
NOTE:
The table power switch must be turned off before disconnecting or
connecting the hand pendant. Once reconnected, turn power on and deploy floor
locks (all four locking feet make firm contact with the floor).
•
With the power switch off, plug the hand pendant into the other hand pendant
receptacle. Turn power switch on. Observe hand pendant buttons illuminate green.
•
Verify the power cord is plugged into a properly grounded AC receptacle.
•
If on battery power, verify proper charge. Observe battery status LED illuminates
green on the hand pendant.
•
Verify power switch is on.
•
Verify floor locks are deployed (all four locking feet make firm contact with the floor);
observe LOCKED WHEN LIT indicator illuminates green on the control panel and
the FLOOR LOCK ON, LED illuminates on the hand pendant.
•
Verify if other hand pendant buttons are functional. If not, replace the hand pendant.