MIZUHOSI 2009
NW0510 Rev. D
28
•
Verify power cord is plugged into a properly grounded AC receptacle.
•
If on battery power, verify proper charge; observe battery status LED illuminates
green.
•
Verify table power switch is on.
•
Activate floor lock override; press and hold FLOOR LOCK OVERRIDE button
(located under control panel). While holding floor lock override control button, press
and hold any hand pendant control button for height up, height down, trendelenburg,
reverse trendelenburg, left lateral roll or right lateral roll. Verify function.
Floor lock button on hand pendant not functioning:
•
Using the hand pendant, press and hold the FLOOR LOCK button. Verify the
LOCKED WHEN LIT green indicator on the control panel or the FLOOR LOCK ON
LED on the hand pendant illuminates.
•
Verify the floor locks are deployed (all four locking feet make firm contact with the
floor). Using the hand pendant, press and hold the FLOOR LOCK button. Try
pressing the FLOOR UNLOCK button and repeat press and hold the FLOOR LOCK
button.
•
Verify power cord is plugged into a properly grounded AC receptacle.
•
If on battery power, verify proper charge; observe battery status LED illuminates
green.
•
Verify the table power switch is on.
•
Activate floor lock override; press and hold the FLOOR LOCK OVERRIDE control
button (located under control panel). While holding the FLOOR LOCK OVERRIDE
control button, press and hold any hand pendant control button for height up, height
down, trendelenburg, reverse trendelenburg, left lateral roll or right lateral roll. Verify
function.
Height up or height down not functioning:
•
Press and hold the hand pendant HEIGHT UP or HEIGHT DOWN control button.
Observe tabletop movement.
•
Verify the hand pendant is properly plugged into a hand pendant receptacle.
Observe hand pendant buttons illuminate green.
NOTE:
The table power switch must be turned off before disconnecting or
connecting the hand pendant. Once reconnected, turn power on and deploy floor
locks (all four locking feet make firm contact with the floor).
•
With the power switch off, plug the hand pendant into the other hand pendant
receptacle. Turn power switch on. Observe hand pendant buttons illuminate green.
•
Verify the power cord is plugged into a properly grounded AC receptacle.
•
If on battery power, verify proper charge. Observe battery status LED illuminates
green.
•
Verify power switch is on.
•
Verify floor locks are deployed (all four locking feet make firm contact with the floor).
Observe LOCKED WHEN LIT illuminates green on the control panel or FLOOR
LOCK ON, LED illuminates on the hand pendant.
•
Verify if other hand pendant buttons are functional. If not, replace the hand pendant.