43
3 Advanced Operations and Useful Functions
No. 99MAM029A
POWER
TOL.
REC/STOP
PRINTER
CLEAR
STAT.
OUT LOG
CANCEL
DATA
FEED
3.6
Other Functions
3.6.1 Timer Input of the Measurement Data
The measurement data can be automatically inputted at a regular interval. The interval
time can be selected from the options below.
0.25 s, 1 s, 5 s, 30 s, 1 min, 30 min, 60 min
Tips
• The measurement data storing (N numbers) and log data storing are feasible for all
timer settings.
• When setting at 0.25 s, only the output with RS-232C is possible but the measurement
data cannot be printed. However, as the measurement data is internally stored, it is
possible for them to become an object for statistical calculation.
• When setting at 1 s, only the output with USB or RS-232C is possible but the measurement
data cannot be printed. However, as the measurement data is internally stored, it is possible
for them to become an object for statistical calculation.
Timer input setting procedure for the measurement data
1
Connect the measuring instrument with Digimatic output or KA Counter to this product
and turn on each power.
2
Enter the timer input mode.
1
Press
FEED
while pressing
PRINTER
.
» "
INTERVAL TIMER
" and the key
name to set the timer input are printed.
3
Start the measurement data input selecting the interval time.
1
Press the key to set the interval time.
» The interval time set and the starting
time/date are printed.
» Timer input starts.
Tips
The buzzer sounds at the time
of measurement data input but it
does not when the interval time
set at 0.25 s.
POWER
TOL.
REC/STOP
PRINTER
CLEAR
STAT.
OUT LOG
CANCEL
DATA
FEED
Set at 5 s
Set at 30 s
Set at 1 min
Set at 30 min
Set at 60 min
Set at 1 s
Interval time
Starting time
and date
Measure-
ment data
Set at 0.25 s