7
Use the minimum cycle parameters below:
Phase
Recirculation
Time
(min:sec)
Water
Temp
Detergent
Type
Pre-wash
2:00
Cold Tap
Water
N/A
Enzyme
wash
4:00
Hot Tap
Water
Enzymatic
Cleaner
Wash
3:00
60°C
Neutral pH
Detergent
Rinse
0:20
Hot Tap
Water
N/A
Thermal
Rinse
1:00
82°C
N/A
Dry
5:00
95°C
N/A
After automated cleaning, proceed to Inspection and
Functional Testing below.
INSPECTION AND FUNCTIONAL
TESTING
• Visually inspect the instrument and check for
damage and wear.
• Cutting edges should be free of nicks and should
have a continuous edge.
• Jaws and teeth should align properly.
• Moveable parts should have smooth movement
without excessive play.
• Locking mechanisms should fasten securely and
close easily.
• Long, thin instruments should be free of bending
and distortion.
REASSEMBLY AFTER CLEANING
To reassemble after cleaning:
1. Insert desired Femoral Rod into Guide Frame.