background image

• Pressing Green 

button will cycle through each of four System 

Modes.  The 

at the top of the display points to the active 

System Mode.

1.OFF – Unit will perform a daily agitation cycle and otherwise 

sit idle.

2.ON – Unit will allow Remote and MANUAL MISTs, but no 

AUTO MISTs.

Section 3

Setting Up the Unit

Using the Controller

• With the display showing day, time 

and level, press Green 

button to 

position the 

at top of display 

pointing to the active System Mode:

• OFF - daily agitation cycle only
• ON – remote and manual, but no 

programmed mist

• AUTO Everyday – program runs 

daily.  This is the usual mode.

• AUTO Custom – program runs on 

days configured in CUSTOM PGM

11. Set System Mode

• Clear area
• Pres MIST button on transmitter
• Unit will FILL, DOSE, insecticide, 

Agitate (AGT) and MIST.

12. Test by Activating 

Remote Mist

• DATA Menu, read AFR (average 

flow rate) value

13. Determine Nozzle 

Flow Rate of Circuit

Section 3

Setting Up the Unit

3.AUTO-EVERYDAY – Misting program runs daily.  This is the 

most commonly used System Mode.

4.AUTO-CUSTOM – Misting program runs only on days set in 

CUSTOM PGM menu.

• Pressing Red MENU/STOP button displays triangular cursor by 

DAY/TIME position on left side of display.  Use 

or 

buttons to 

cycle through menu structure.  Use Green 

button to select 

menu item and view or change data element within that item.

• Within a menu item, the convention is that the flashing data 

element can be changed with the 

or 

buttons.  Move to the 

next data element with the Green 

button.

• Exit the menu item by pressing the MENU button.  The unit will 

revert to previous System Mode in 3 minutes if the MENU button 
is not pressed.

• Clear any error code by pressing the Red MENU/Stop button for 

5 seconds.  The System Mode will return to its previous state 
(usually AUTO Everyday).

• To reset the controller to factory settings, unplug the unit, and 

hold down MENU and the 

buttons while plugging the unit 

back in.  Mist Cycles, remote duration and number of nozzles 
must be reentered.

• Secure key in a safe location.

14. Close Lid and Lock

flow rate) value

• Enter that value as the nozzle flow 

rate, NFR, to be used in system 
error calculations

• DATA Menu, NFR. Press Green 

button for 5 seconds, then use the 

or 

buttons to adjust value to be 

the same as read in AFR.

Flow Rate of Circuit

Отзывы: