transient/burst
IEC 61000-4-4
input/output
lines
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
Not Applicable
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5 %
U
T
(>95 % dip in
U
T)
for 0,5 cycle
40 %
U
T
(60 % dip in
U
T)
for 5 cycles
70 %
U
T
(30 % dip in
U
T)
for 25 cycles
<5 %
U
T
(>95 % dip in
U
T)
for 5 sec
Not Applicable
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
30 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE
U
T is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration –
electromagnetic immunity
The
Smart One
is intended for use in the electromagnetic environment specified below. The
customer
or the user of the
Smart One
should assure that it is used in such an environment.
Portable and mobile RF communications equipment
should be used no closer to any part of the
Smart
One
, including cables, than the recommended