USER MANUAL
SMALL-SIZE INFRARED GAS SENSOR MIPEX-02-3-RX-1.1 G
ESAT.100200.02.02 UM
September 23, 2021
Page 6 of 41
THE SOLE PROPERTY OF SIA MIPEX. ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS PROHIBITED.
3. TECHNICAL SPECIFICATIONS
Table 2. General specifications (for available options see Appendix A)
Gas sampling method
Diffusion
Operating principle
Non-Dispersive Infra-Red (NDIR)
Target gas
1
CO
2
Operating,
storage and
transportation
conditions:
Relative humidity, %
up to 98
Atmospheric pressure, kPa
80…120
Temperature
2
, °C
-55...+60
Temperature range, °C
-10…+40
-20…+50
Overall dimensions, mm
⌀
20.3×16.6
(without pins)
Pins length, mm
4.6
Warm-up time, sec
≤ 120
Weight, g
17.2
Housing material
Stainless steel
Life time expectancy
3
(not less than), years
10
Shelf life time (not less than), years
8
IP rating
20 (without dust filter)
54 (if a dust filter provided by SIA MIPEX is applied)
1
See Appendix A for details.
2
Term “operating temperature” refers to ambient temperature, at which sensor operates, but its readings
variability stated in Table 3 is provided only in specified temperature range (see Table 5 and Table 6).
3
To provide metrological properties during sensor lifetime, zeroing and span calibration should be
performed periodically, at least once in 30 months (see Appendix D.1).