USER MANUAL
SMALL-SIZE INFRARED GAS SENSOR MIPEX-02-3-RX-1.1 G
ESAT.100200.02.02 UM
September 23, 2021
Page 33 of 41
THE SOLE PROPERTY OF SIA MIPEX. ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS PROHIBITED.
Appendix C.2.5. Commands for configuring net address
Request syntax
USER
mode
OEM
mode
Response description
Command description
!**
no
yes
4 bytes in ASCII format: “0D”
symbol, sensor’s hexadeci-
mal net address in range
00…FF and “0D” symbol.
Returns sensor’s net address
%AABB
no
yes
Three bytes in ASCII format:
sensor’s new hexadecimal
net address and “0D” symbol.
Intended to assign a new net
address
AA
to sensor with
address
BB
.
NETOFF
no
yes
NETOFF OK
or
NETOFF FAULT
Cancels storing of sensor’s net
address in sensor memory.
After sensor receives this
command and sensor power
resets, sensor’s net address
becomes “00”.
NETON
no
yes
NETON OK
or
NETON FAULT
Allows assigning and storing
sensor’s net address by using
<%AABB> command.