2.  DESCRIPTION 

MIPEX-02  sensor  is  a  small-size  gas  sensor  that  operates  on  the  NDIR  (non-dispersive 

infrared radiation) principle to detect and determine the concentration of carbon dioxide (CO2) in a 
areas. The sensor is used in: food industry, agricultural industry and production technological cycles. 
The sensor is used as a OEM component for portable and fixed gas detectors. Has a long life time. 
The sensor's low power consumption allows it to be used in battery-powered systems, including IoT 
applications.  

Sensor has several modifications for specific needs. They differ by housing types, photodiode 

and LED, calibration gases, interface types, measuring ranges etc. (see Appendix A).  

Sensor is a smart integrated system and includes mirror optical system, photodiode and LED, 

signal amplifiers, microcontroller, current driver of the infrared LED, UART interface signal generator 
and  supply  forming  voltage  unit.  Sensor  microcontroller  performs  storage  of  unique  sensor 
calibration constants, processing of measurement results and concentration of measured gas, and 
information exchange. 

Sensor operating principle is based on NDIR technology, i.e. on selective absorption of infrared 

radiation by carbon dioxide molecules in its spectral absorption band. 

Differential  dual  wavelength  method  allows  eliminating  of  water  vapor,  optical  elements 

contamination and other non-selective hindrances influence. 

Target gases available for MIPEX-02 are listed in Table 1. 

 

Table 1. Target gases for MIPEX-02 sensors 

Target gas 

Analytical tasks 

description 

Optical elements 

code (see 

Appendix A) 

Spectral 

characteristic 

maximum 

Notes 

CO

2

 

Analyzing  the  atmos-
phere  (including  of  in-
dustrial  facilities)  con-
taining carbon dioxide. 

4.2 

µ

Sensor is selective to 
other gases. 

Содержание 02-3-RX-1.1 G

Страница 1: ...www mipex tech com ESAT 100200 02 02 UM September 23 2021 Page 1 of 41 SMALL SIZE INFRARED GAS SENSOR MIPEX 02 3 RX 1 1 G Non Ex version USER MANUAL ESAT 100200 02 02 UM...

Страница 2: ...PERMISSION OF SIA MIPEX IS PROHIBITED INFORMATION CONTAINED IN THIS DOCUMENT IS THE SOLE PROPERTY OF SIA MIPEX ANY REPRODUCTION IN PARTS OR AS A WHOLE WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS P...

Страница 3: ...ortation 15 7 Warranty 16 8 Contacts 17 Appendix A Sensor types and characteristics 18 Appendix B Connection diagram 23 Appendix C UART communication protocol 24 Appendix C 1 General information 24 Ap...

Страница 4: ...sor SIA MIPEX reserves the right to make changes to this manual 1 1 List of abbreviations ADC Analog to Digital Converter LED Light Emitting Diode CGM Control Gas Mixture NDIR Non Dispersive Infra Red...

Страница 5: ...tc see Appendix A Sensor is a smart integrated system and includes mirror optical system photodiode and LED signal amplifiers microcontroller current driver of the infrared LED UART interface signal g...

Страница 6: ...5 60 Temperature range C 10 40 20 50 Overall dimensions mm 20 3 16 6 without pins Pins length mm 4 6 Warm up time sec 120 Weight g 17 2 Housing material Stainless steel Life time expectancy3 not less...

Страница 7: ...on whichever is greater Response time T90 without dust filter sec 30 Response time T90 when dust filter provided by SIA MIPEX is installed see Appendix E sec 60 Table 4 Electrical specifications and s...

Страница 8: ...sensor reflecting cover or on sensor side surface 2 MPa on sensor upper edge 100 MPa Sensor updates information about concentration every 1 28 0 065 second Sending any command more often than one time...

Страница 9: ...thout power supply after transportation as well as after dust filter is applied see Appendix D 1 It is required to perform span calibration at least once in 30 months 5 1 Preparation If sensor has bee...

Страница 10: ...tion Pulse repetition period is 80 20 ms depends on ambient temperature During operation maximum surge current is 17 mA for 20 microseconds see Fig 2 Average current consumption is not more than 330 A...

Страница 11: ...100200 02 02 UM September 23 2021 Page 11 of 41 THE SOLE PROPERTY OF SIA MIPEX ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS PROHIBITED Fig 2 Typical current consumption waveform Fig...

Страница 12: ...02 02 UM September 23 2021 Page 12 of 41 THE SOLE PROPERTY OF SIA MIPEX ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS PROHIBITED Fig 4 Typical barrier voltage drop waveform 3 3 V in...

Страница 13: ...command via UART interface for DATAE2 command description see Appendix C 2 1 for status bytes description and recommendations see Table 12 Sensor span calibration and zeroing should be performed next...

Страница 14: ...nnection diagram of multiple sensors to single UART line Resistors are calculated as follows 1 Determine the quantity of sensors N that should be connected to single UART line 2 Determine forward volt...

Страница 15: ...Transportation can be done by all means of transportation in covered vehicles as well as in a heated pressurized plane compartments in accordance with the rules of cargoes transportation actual for th...

Страница 16: ...corrosion oxidation ingression of any foreign objects materials liquids insects if sensor was damaged by using connection circuit that does not comply with requirements stated in this UM if damage is...

Страница 17: ...1 Page 17 of 41 THE SOLE PROPERTY OF SIA MIPEX ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS PROHIBITED 8 CONTACTS MIPEX TECHNOLOGY SIA MIPEX 2B Valkas iela Daugavpils Latvia 5417 Te...

Страница 18: ...ON OF SIA MIPEX IS PROHIBITED APPENDIX A SENSOR TYPES AND CHARACTERISTICS See Table 5 for details See Table 6 for details MIPEX 02 B RX D 1 G Interface 1 UART 4 pins Housing type 1 sensor housing made...

Страница 19: ...IA MIPEX ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS PROHIBITED Table 5 Metrological properties of sensor RX code Part number Target gas Calibration gas Meas range vol Temp range C...

Страница 20: ...5 of readings whichever is greater within the range of 20 25 C 0 1 vol or 40 of readings whichever is greater at 100 kPa tested at 80 kPa 100 kPa 120 kPa 0 1 vol or 15 of readings whichever is greater...

Страница 21: ...R MIPEX 02 3 RX 1 1 G ESAT 100200 02 02 UM September 23 2021 Page 21 of 41 THE SOLE PROPERTY OF SIA MIPEX ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS PROHIBITED Fig 7 MIPEX 02 3 RX...

Страница 22: ...2 3 RX 1 1 G ESAT 100200 02 02 UM September 23 2021 Page 22 of 41 THE SOLE PROPERTY OF SIA MIPEX ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS PROHIBITED Table 7 Sensor pinout Pin Pu...

Страница 23: ...0 02 02 UM September 23 2021 Page 23 of 41 THE SOLE PROPERTY OF SIA MIPEX ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF SIA MIPEX IS PROHIBITED APPENDIX B CONNECTION DIAGRAM Fig 8 Connection diag...

Страница 24: ...st responses command ends with carriage return symbol hex 0D words and or values in a response on most commands are separated by space symbol hex 20 or in some cases by tabulation hex 09 all symbols i...

Страница 25: ...requesting measured data and diagnosing see Appendix C 2 1 2 commands for controlling operating mode see Appendix C 2 2 3 commands for requesting factory settings and properties see Appendix C 2 3 4...

Страница 26: ...CII format tabulation symbol 5 bytes in ASCII format and 0D symbol Returns sensor readings of scaled concentration C1 current ambient temperature in Celsius degrees and status word see Table 16 CFS ye...

Страница 27: ...ata C1 hex 0D Table 10 Structure of response to DATAE command Command DATAE hex 44 41 54 41 45 0D Byte number 1 2 3 4 5 Byte data7 C1H C1L StatusByte CRC hex 0D Table 11 Structure of response to DATAE...

Страница 28: ...ature mode temperature changes faster than by 0 6 C min Do not perform span calibration or zeroing 5 5 22 Dynamic temperature mode temperature changes faster than by 2 C min Do not perform span calibr...

Страница 29: ...r switching com mands level from USER to OEM XXXX stands for old password and YYYY stands for new one New password should contain only digits 0 9 in ASCII format UART yes yes USER or OEM Returns activ...

Страница 30: ...value incre mented to 10000 for example to write the scale coefficient of 0 009 send CALB2 00090 This com mand can be used to recalibrate sen sor if scale coefficient is known The scale coefficients f...

Страница 31: ...check sensor readings In case if sensor does not comply to its metrological specifications perform span calibration see Appendix D 1 SETC XXXXX yes yes SETC XXXXX OK or SETC XXXXX FAULT Intended for d...

Страница 32: ...symbol Returns a number contained in cell XX 00 09 of sensor memory user section USERDATAXX YYYYY no yes USERDATAXX YYYYY OK Writes a number YYYYY 00000 99999 to cell XX 00 09 of sensor memory user se...

Страница 33: ...SCII format 0D symbol sensor s hexadeci mal net address in range 00 FF and 0D symbol Returns sensor s net address AABB no yes Three bytes in ASCII format sensor s new hexadecimal net address and 0D sy...

Страница 34: ...being carried out avoid excessive pressure on sensor housing 98 humidity pressure of flowing gas less than 99 kPa and more than 103 kPa temperature changes faster than by 0 6 C min dust ingress if dus...

Страница 35: ...A ensure the following conditions 4 1 entered concentration value must not deviate from the current readings of more than 20 times and also must not be less than 11 C 20 C C 0 05 C 10 4 2 CGM must con...

Страница 36: ...MIPEX IS PROHIBITED Table 14 CGMs used for CO2 span calibration CGM number according to text Components composition Content of carbon dioxide vol Permissible deviation limits vol Limits of permissibl...

Страница 37: ...example of procedure for temperature range of 40 60 C 1 Fill sensor atmosphere with nitrogen 2 Power on the sensor 3 1 minute after send ZERO0 command This command erases previously defined temperatur...

Страница 38: ...ic membrane Vladipor MFFC TC 6 55 221 1413 2004 On the filter backside there is an adhesive layer 3M Double Linered Laminating Adhesive 7952 in hatched areas in Fig 12 Fig 12 Top filter dimensions Per...

Страница 39: ...support Contacting support send log file with periodic responses to command F which is intended exclusively for detailed diagnostics It returns comprehensive information on sensor readings and its st...

Страница 40: ...IA MIPEX IS PROHIBITED Table 15 Structure of response to F command Command F hex 46 0D response Byte number Byte data 1 Hex 0E 2 6 T 7 Hex 09 8 12 St 13 Hex 09 14 18 Us 19 Hex 09 20 24 Uref 25 Hex 09...

Страница 41: ...re on optical elements Dry sensor After drying if this status lasts longer than for 20 minutes at stable conditions replace sensor 31 8 Zero shifts to negative value Possibility of condensed moisture...

Отзывы: