background image

13

TROUBLESHOOTING

 Always disconnect power to the water purifier before performing any service or maintenance. 

IMPORTANT: This unit is to be serviced ONLY by qualified, and appropriately licensed, personnel.

Table 3 - Troubleshooting

Problem

Possible Cause

Corrective Action

Purifier not operating

No electrical power

Verify that the purifier is connected to a live 
power source.

Water leaking from purifier

Cracked or broken quartz sleeve

Shut down purifier, drain, and replace quartz sleeve. 
See 

“Replacement of Broken Quartz Sleeve”

 in the

 

“Maintenance”

 section.

Quartz sleeve sealing O-Ring(s) 
worn, damaged

Shut down purifier, drain, and  remove static gland 
nut, replace sealing O-Ring. See 

“Quartz Sleeve 

Installation, Cleaning or Replacement”

 in the

 

“Maintenance”

 section.

Poor, or loose, connections  
or fittings

Tighten suspect connection or fitting; or shut down 
purifier, drain, and remove fitting or connection. Clean 
threads; reapply thread sealing tape and reinstall.

Poor purifier disinfection 
performance

Quartz sleeve fouled

Clean quartz sleeve, see “

Quartz Sleeve Installation, 

Cleaning or Replacement

” Section.

Germicidal lamp output 
depreciating

Replace lamp, as it nears its end of life 

(EOL)

See “

Lamp Installation or Replacement

” in the 

Maintenance

” section.

Germicidal lamp not functioning

Replace lamp. See “

Lamp Installation or 

Replacement

” in the “

Maintenance

” section.

Low input voltage

Verify input voltage to purifier.

Change in water quality

Have water tested to confirm that it does not exceed 
maximum recommended concentration levels for use 
with this purifier.

b

allast

 n

o

. 10-0508b 

and

 10-0504b 

Have

 l

amP

-

out

 P

roteCtion

If power is applied to the ballast without a lamp connected to the circuit or if the lamp fails, the ballast’s lamp-out protection 
will prevent the ballast from operating, and a red LED light will illuminate and audio alarm will sound. 

To reset: disconnect the 

ballast from power and reconnect or replace the lamp.

 The ballast must remain disconnected from power for (2) minutes to 

reset. After 2 minutes restore input power, ballast should resume proper operation.

 

The equipment that this ballast is used with requires connection to a known good ground. It is the User’s 

responsibility to ensure that their equipment is properly grounded in compliance with the National Electrical Code and any 
applicable local codes. Equipment wiring and hook up should be performed by a qualified, licensed electrician. Failure to comply 
with this requirement could result in serious injury or death. 

Содержание MIN-6

Страница 1: ...2017 Atlantic Ultraviolet Corporation ASSEMBLED IN THE USA Installation Operation Maintenance Read and Follow All Safety Instructions Save These Instructions Models Ballast 120v 50 60 Hz Ballast 230v 50 60Hz MIN 6 10 0508B 10 0504B MIN 9 10 0508B 10 0504B Owner s Manual Owner s Manual ULTRAVIOLET WATER PURIFIERS ...

Страница 2: ...lieved to be correct However no guarantee or warranty of any kind expressed or implied is made with respect to the information contained herein Specifications and information are subject to change without notice SAFETY WARNINGS 3 SAFETY INSTRUCTIONS 3 SAFETY LABELING 3 CAUTION 3 PRODUCT APPLICATION 4 Construction 4 Principle of Operation 4 Limitation of Use 4 Water Quality 4 INSTALLATION 5 Locatio...

Страница 3: ...ot avoided COULD result in death or serious injury This symbol pictorial is used to identify the need to wear approved ultraviolet blocking eyewear This symbol pictorial is used to identify the need to wear approved ultraviolet blocking face shield This symbol pictorial is used to identify components which must not be disposed of in trash Danger indicates an IMMINENTLY hazardous situation which if...

Страница 4: ...z sleeve and the chamber wall Suspended microorganisms are exposed to the ultraviolet rays emitted by the germicidal lamp The LED indicator light located on the ballast provides visual indication of germicidal lamp operation Water leaving the purifier is instantly ready for use no further contact time is required Limitation of Use The water purifier is intended for the use with visually clear wate...

Страница 5: ...will not result in damage to the area surrounding the water purifier For added protection it is recommended that a suitable drain pan be installed under the purifier The drain pan must be plumbed to an adequate free flowing drain to prevent water damage in event of a leak There are numerous leak detection flood stop devices available on the market today designed to stop flow of water reducing the ...

Страница 6: ...ble 2 MIN 6 and MIN 9 Dimensional Data Model A B C D E F Inlet Outlet MIN 6 14 3 4 22 17 32 5 3 4 4 1 4 22 5 3 8 3 4 m NPT MIN 9 21 7 8 29 17 32 5 3 4 4 1 4 29 5 3 8 3 4 m NPT All specifications dimensional data etc are approximate and subject to change without notice ...

Страница 7: ... entering the water purifier Particulates may cause damage to the quartz sleeve as well as interfere with the purifier s ability to disinfect the water The sediment filter may also help to reduce the amount of routine cleanings of the quartz sleeve 6 Shut off valves should be installed on both the inlet and outlet sides of the water purifier The use of bypass valves is not recommended The shut off...

Страница 8: ... Safety Sensor alerts the user to any malfunction detected 3 PromateTM Time Delay Mechanism Provides a 2 minute warm up period during which the ultraviolet lamp achieves its full germicidal output before the water is allowed to flow through the water purifier The time delay mechanism is used in conjunction with and is electrically connected to the Solenoid Valve 4 PromateTM Solenoid Valve Operated...

Страница 9: ...rovides visual indication of germicidal lamp operation This provides an indication of lamp operation and does not indicate the level of ultraviolet intensity or transmission through the water Ballasts have a built in audio alarm which will indicate lamp failure On purifiers so equipped the SENTRYTM Safety Sensor provides visual indication of ballast and ultraviolet lamp operation 2 To ensure prope...

Страница 10: ...hamber Be sure to withdraw lamp straight out without angling until completely clear of quartz sleeve Lamp and quartz sleeve are easily damaged Exercise care when handling 5 Reinstall lamp in reverse order Germicidal ultraviolet rays are harmful to eyes and skin Do not restore power to water purifier until lamp and both EASY OFFTM End Caps have been properly reinstalled Step 1 Disconnect Power to W...

Страница 11: ...e quartz sleeve from chamber NOTE It is advisable to support the quartz sleeve on the opposite end with your finger so that it does not drop to the bottom of the chamber as it slides into the chamber 7 Once the quartz sleeve is removed clean with alcohol or a mild non abrasive detergent Stubborn stains usually can be removed with a dilute hydrochloric acid NOTE Follow all manufacturer s instructio...

Страница 12: ... quartz sleeve parallel to the chamber angling the quartz sleeve in any direction could result in the breakage of the quartz sleeve 11 As the quartz sleeve nears the far end of the chamber support the quartz sleeve by inserting your finger through the far end gland fitting and into the quartz sleeve This will minimize the possibility of breaking the quartz sleeve as it passes through the gland fit...

Страница 13: ...amp as it nears its end of life EOL See Lamp Installation or Replacement in the Maintenance section Germicidal lamp not functioning Replace lamp See Lamp Installation or Replacement in the Maintenance section Low input voltage Verify input voltage to purifier Change in water quality Have water tested to confirm that it does not exceed maximum recommended concentration levels for use with this puri...

Страница 14: ...tional WARNINGS 12volt and 24volt Direct Current DC ballasts and units Many Individuals regard ballasts and units powered by 12volt and 24volt Direct Current DC to be relatively benign from a safety standpoint However a DC ballast in order to operate an ultraviolet discharge lamp must produce an output as high as 600volts This voltage is high enough to produce a very serious or even lethal electri...

Страница 15: ... 1 10 0508B 10 0508B 1B SteadfastTM Lamp Socket SureliteTM Ballast 230v 50 60Hz 1 10 0504B 10 0504B 2 Rubber O Ring Static Gland 2 00 1108B 00 1108B 3 STER L RAY Lamp 1 05 1370 R 05 0097A 4 Static Gland Nut 2 25 1492D 25 1492D 5 Teflon Washer 2 25 1235A 25 1235A 6 Rubber O Ring Quartz Sleeve 2 00 1238A 00 1238A 7 CRYSTAL CLEARTM Quartz Sleeve 1 15 1112A 15 1006A3 8 EASY OFFTM End Cap no hole 1 25 ...

Страница 16: ... will be free from defects in material and workmanship for a period of one year from the date of shipment thereof or the product s total rated life whichever first occurs Within the warranty period we shall repair or replace such products which are returned to us with shipping charges prepaid and which are determined by us to be defective This warranty will not apply to any product which has been ...

Отзывы: