38
6-2-3. Connecting Grounding Wire
When an amplifier unit is connected to a case, the protective grounding terminal, case protective
grounding terminal of batch powr supply connector are connected and have the same potential.
The grounding wire should have AWG16 wire material and connected with a screw.
Always connect the grounding wire to gound.
6-2-4. Caution on Rack Mount case
A rack mount case is a case that is used for accommodation a case to 19 inch rack.
The rails on the left and right should be placed on the rails of the rack, and then fix the case by
using four attaching holes. When using several cases, install a fan unit between cases in order to
release heat and maintain amplifier accuracy.
Avoid placing a rack mount case on a desk or floor as the rack mount case is noprovided
with rubber legs.