20
3-6.
Case
3-6-1. Names of case Elements
Fig. 3-6
Use standard AC power cord (47326) that is attached to the case as a standard for 100 VAC power
supply.
Use optional DC power cord (47339) for DC power supply.
3-6-2. How to use case
(1)
Power suppy cable connection
When using battery (12 VDC, or 10 VDC to 30 VDC) for power supply with amplifier units being
installed in a case, take into account the voltage drop caused by the DC power cord length and wire
diameter. When using several channels or a long poer supply cord, a voltage drop occurs. This
voltage drop may cause the power supply voltage to be lower than the allowable powr supply voltage
of 10 V at DC power input connector.
For example, the DC power supply cord (47229) has a core area of 1.25 mm
2
. If eight amplifiers are
mounted, a current of 3.2 A (0.4 A x 8) flows, and a voltage drop of 0.5 V will occur if the cord is
extended to 10 m. If a 10-m cable with 0.75 mm
2
is used, a voltage drop of 1.5 V will occur. For use
under such conditions, the power supply should be provided with the voltage drop taken into
account, or the wire diameter or cord length of the power cord should be modified.
(2)
Synchronization between units
When two or more amplifier units are installed in a case, unit synchronization is maintained by
theinternal wiring inside the case and synchronization signal. One amplifier should be set to be the
master, and others should be set to be slaves. To set the unit to be the master, set synchronization
switch 2 on the rear face panel to INT; to set to the slaves, set the switch to EXT. Note that only
amplifier units that have the same bridge power frequency can be installed in a case. It is
impossible to share a common case or synchronize between cases that have amplifiers installed with
different bridge power frequencies. Take notice that the bridge power frequency of DSA-631 is 5
kHz.
Do not set two or more units to INT. Otherwise correct measurement cannot be made, or this may cause
a defect in amplifier units.
AC Power input
AC Power fuse
DC Power fuse
Ground terminal
DC Power input
Interface Connector
CAUTION
Содержание DSA-631
Страница 1: ...Dynamic Strain Amplifier DSA 631 INSTRUCTION MANUAL EN294 1320 E ...
Страница 2: ......
Страница 58: ...48 9 3 Outline dimensions 9 3 1 Single unit Unit mm ...
Страница 59: ...49 9 3 2 Panel cut size Square hole Unit mm ...
Страница 61: ...51 9 3 4 Rack mount case Rail for rack Long hole Unit mm ...
Страница 62: ...52 9 3 5 Bridge box FA410 072 model Unit mm ...
Страница 63: ......