11
Number Name
Function
4
Output connector 1 (OUTPUT1)
The output voltage and current are
±
10 V and
±
5 mA,
respectively. This connector can be connected to a
recorder (e.g. Thermal-dot recorder or data acquisition
devices ) or A/D converter, which accept voltage signals.
5
Output connector 2 (OUTPUT2)
The output voltage and current are
±
10 V and
±
5 mA,
respectively. The output level can be controlled from 10
V to 1 V with OUTPUT2 level adjustment volume (18)
on the front panel.
Note) If output of current from 4 mA to 20 mA is
required, modification from the voltage output to the
current output is available with extra costs. For details,
contact with us or sales representatives.
6
Interface connector (I/F)
This connector is used to connect an amplifier unit and
case electrically. The pin layout is shown as follows.
Other than DC power voltage supply, it is possible to
perform keylock, auto-balancing, calibration value
application, and synchronous signal output.
A connection cable is available separately.
Amplifier, Interface connector (from rear side)
①
+
CAL
②
-
CAL
③
BAL
④
OSC
⑤
GND
⑥
KEYLOCK
⑦
GND
⑧
DC
+
⑨
DC
-
7
Protective grounding terminal
If 3-pin power cord cannot be used for grounding, use
this terminal for grounding.
8
Power supply connector
This is the connector to be connected to the AC power
cable. The AC power supply block of the amplifier unit
have a withstand voltage of 1.5 kVAC/minute against
input, output, and case.
Note) When using a power supply of 100 VAC, use AC
power supply cord 47326. When using 110 VAC or
higher, use optional AC power supply cord 200 V
(0311-5112). When using 180 VAC to 264 VAC, also use
optional AC power cord (0311-5112).
** If this amplifier is used in Japan domestic, the
voltage exceeding power supply voltage AC125 V
cannot be supplied from the limitation of “Electrical
Appliances and Material Safety Law”.
Table 2-2 Rear Panel : Names and Functions
Содержание DSA-631
Страница 1: ...Dynamic Strain Amplifier DSA 631 INSTRUCTION MANUAL EN294 1320 E ...
Страница 2: ......
Страница 58: ...48 9 3 Outline dimensions 9 3 1 Single unit Unit mm ...
Страница 59: ...49 9 3 2 Panel cut size Square hole Unit mm ...
Страница 61: ...51 9 3 4 Rack mount case Rail for rack Long hole Unit mm ...
Страница 62: ...52 9 3 5 Bridge box FA410 072 model Unit mm ...
Страница 63: ......