4
4
.
Setting of CC-Link connecting
Please set the following when you use CC-LINK interface.
4-1.
Changeover of CC-Link communication setting change screen
Change from the standard measurement mode to the function mode by the following operations.
1) Touch the MENU key.
2) Touch the NEXT key.
3)
Touch the OPTION key.
4)
Touch the CC-LINK setting key.
4-2.
Items of CC-Link setting
(
1
)
The number of occupied stations.
Number of occupied stations is set.
Selectable from [1 station], [2stations], [4 stations].
Default has set as [4 stations].
(
2
)
Setting of the stations
Execute the setting of the station No.
When 1 station is occupied : selectable from station No.01 to 64.
When 2 stations are occupied : selectable from station No.01 to 63
When 4 stations are occupied : selectable from station No.01 to 61.
The occupied station of this instrument is 1., 2, 4 stations.
※
When 2 stations are occupied,and station No. is set as “01”,01 to 02 stations are occupied.
※
When 4 stations are occupied,and station No. is set as “01”,01 to 04 stations are occupied.
Therefore, the station number must not overlap.
Default has selected [01].
(
3
)
Setting of baud rate
Communication speed (unit: bps) is set.
Selectable from [156k], [625k], [2.5M], [5M] or [10M].
Default has set as [156k].
•
Setting changes for occupied stations No. is corresponding to this software after ROM Ver.1.300.Before ROM Ver.1.200 is 4
station occupied station No.