2
3.
Connecting method
3
−
1. Layout of the terminal boards
There is the terminal boards, which has 10 points of terminals.
Layout of terminal boards are shown in the following figure.
:
●
Terminals(CN1)
Name of
terminals
Description
Applications
1
A
Bridge power supply(
+
)
2
B
Amplifier input(
−
)
3
C
Bridge power supply(
−
)
Strain gage applied transducer
4
D
Amplifier input(
+
)
5
E
Shield
6
V
+
Voltage output terminal(
+
)
Voltage output
7
V
−
Voltage output terminal(
−
)
Voltage output
8
F.G.
Grounding terminal
9
DC24 V(
+
)
Power supply input terminal(
+
)
Power supply
10
DC0 V(
−
)
Power supply input terminal(
−
)
pp y
●
“V
−
” of voltage output terminal and “DC0 V (
−
)” of power supply input terminal are
insulated.
●
“C” of bridge power supply (
−
) and “V
−
” of voltage output terminal are connected
internally.
3
−
2. Note on connection
Warning
●
In case of connection with the instrument, keep strictly to the following items. If
neglected, it may cause an unexpected failure or a damage to the instrument.
D
Be sure to set the power supply to OFF, when the connection is made.
D
Since the terminal boards at front of the instrument is made of resin, take care not to drop
it down or not to apply strong impact.
D
The electric wire which suits the terminal block of this unit is 0.08 mm
2
/AWG28
〜
2.5
mm
2
/AWG12.
D
The electric wire bare length is 5 mm to 6 mm.
D
Connecting cable with the instrument should be away from the noise source such as power
supply line and/or I/O line for control and so on as far as possible.
D
Conduit wiring should be the type of exclusive one, and avoid using with another line
together.
D
All of the connections should be executed securely by referring to the Instruction manual
for the instrument.
Содержание CSA-521
Страница 1: ...Transmitter CSA 521 Instruction Manual EN294 1145 G ...
Страница 2: ......
Страница 10: ......
Страница 31: ......