29
Remove the connecting cable for strain gage applied transducer from the terminals
Set the connecting range to DC
・
V for the measuring instrument such as
tester and so on
1
Re
−
connect the connecting cable of the strain gage applied transducer with the
terminals(Refer to the section 4)
2
Checking the voltage between the terminal B and D on the terminals, apply the
load on strain gage applied transducer
Set the connecting range to DC
・
mV on the measuring instrument such as
tester and so on
3
Contact
us
YES
NO
YES
YES
NO
NO
There may have fears that
connected strain gage applied
transducer has broken or signal
route(signal wire) has cut off
Inform
us
about
the
contents
of
failure
and
situation
at
site
in
details
The voltage between
A and C is just like the setting
of DIP2(refer to 6
−
3)
The voltage between
B and D varies
Make calibration again according to the
procedures in the section 5
Same condition
Start of measurement