3.1.4
Checking the equipment supplied
-
1 floor platform
-
Safety instructions and QR code for access to the complete documentation
3.1.5
Leveling the weighing platform
To achieve reproducible weighing results at all times, the weighing platform must be set
up to be precisely horizontal.
Therefore the weighing platform must be re-leveled every time it is moved to a different
location.
Leveling the weighing platform
.
Unpack the device and check it for visible external damage.
If there is damage, follow the instructions in the chapter "Safety check".
.
Keep the original packaging in case the device needs to be returned. Remove all
cables before sending.
.
Use the adjustable feet to align the weighing platform so that the air bubble of the
level indicator is in the center of the circle.
.
Check that all four of the adjustable feet are touching the surface.
The weight of the platform must be spread equally across the adjustable feet.
.
Adjust the adjustable feet: Retract the adjustable feet (clockwise) in order to lift the
scale. Extend the adjustable feet (counter-clockwise) in order to lower the scale.
3 Device installation
Floor platform Puro®
Minebea Intec
EN-7