background image

NOTICE

Welding or lightning strike current lowing through the cell can damage it.

All electrical welding on the weighing system must be inished before mounting the load

cells.

NOTICE

Corruption of measurement results!

The elastomer support must not be used to absorb transverse forces.

4.3.2

Installing the mounting kit and inserting the load cell

Note:

How to mount the vessel in the frame structure is described below.
If elastomer supports are used, they must be provided on all load cells.
The work steps must be carried out on all support points of the weighing object.

Procedure:

1. At the supporting point (elastomer support / vessel frame), make the borehole for the

elastomer support (1) in the vessel frame.

2. Mount the load cell (5) with spacer (4) on the lower mounting plate (3). The property

classes and tightening torques of the screws (2) must be observed (see Chapter

4.2

.

3. Screw in the elastomer support (1).
4. Mark the boreholes for the lower mounting plate (3) on the foundation and make the

holes.

5. Fix the mounting kit to the foundation. The property classes and tightening torques

of the screws must be observed; see Chapter

4.2

.

4 Installation

Mounting kit with elastomer support PR 92/00N

Minebea Intec

EN-9

Содержание PR 92/00N

Страница 1: ...upport PR 92 00N Installation manual Translation of the original installation manual 9499 053 92000 Edition 7 1 0 10 17 2019 Minebea Intec GmbH Meiendorfer Str 205 A 22145 Hamburg Germany Phone 49 40...

Страница 2: ...iied personnel In correspondence concerning this product the type name and release number serial number as well as all license numbers relating to the product have to be cited Note This document is pa...

Страница 3: ...4 2 3 Initial inspection 4 2 4 Before operational startup 4 3 Speciications 5 3 1 Equipment supplied 5 3 1 1 Mounting kit PR 92 00N 5 3 2 Dimensions 6 3 3 Technical data 7 4 Installation 8 4 1 Prior...

Страница 4: ...measures for preventing hazards are not followed DANGER Warning of personal injury DANGER indicates death or severe irreversible personal injury which will occur if the corresponding safety measures a...

Страница 5: ...ironment may occur if appropriate safety measures are not observed Take the corresponding safety precautions Note User tips useful information and notes 1 6 Hotline Phone 49 40 67960 444 Fax 49 40 679...

Страница 6: ...over or being shifted thus eliminating danger to people animals or goods even in the case of a break in a load cell or mounting element Installation and repair work must only be carried out by expert...

Страница 7: ...plied 3 1 1 Mounting kit PR 92 00N No Description 1 Elastomer support 2 Lower plate 3 Mounting bolt 2 4 Washer 2 5 Spacer plate 6 Quick guide not shown 3 Speciications Mounting kit with elastomer supp...

Страница 8: ...sions PR 92 00N All dimensions in mm Ensure the spacer plate is in the correct position Load cell MP 58 126 4 mm MP 79 127 3 mm Mounting kit with elastomer support PR 92 00N 3 Speciications EN 6 Mineb...

Страница 9: ...22 kN Permissible horizontal displacement 2 5 mm at 1 kN Permissible temperature range 20 C 65 C Material Elastomer support Lower plate Spacer plate Natural rubber Steel galvanized chromated and seale...

Страница 10: ...ampening or for temperature insulation a load compensating plate must be provided between this soft iller layer and the mounting kit to ensure even load application into the mounting kit The design of...

Страница 11: ...ps must be carried out on all support points of the weighing object Procedure 1 At the supporting point elastomer support vessel frame make the borehole for the elastomer support 1 in the vessel frame...

Страница 12: ...asured object due to thermal expansion vibration etc can be utilized without reducing the measuring accuracy only if the load cell has been installed exactly To avoid force shunts all incoming and out...

Страница 13: ...bottom and around the mounting kit NOTICE Some cleaning agents may not be compatible with the mounting kit material When using cleaning agents ensure that their compatibility with the mounting kit mat...

Страница 14: ...municipal waste so that it can then be recycled Before disposing of or scrapping the product any batteries should be removed and taken to a suitable collection point Please see our T Cs for further i...

Страница 15: ......

Страница 16: ...Published by Minebea Intec GmbH Meiendorfer Strasse 205 A 22145 Hamburg Germany Phone 49 40 67960 303 Email info minebea intec com www minebea intec com...

Отзывы: